SEDECAL Soltus 500 (REF 10501)Recalls & reported problems
Sedecal · Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08436046002456), received between Jul 22, 2022 and Dec 15, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Unintended System Motion” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Aug 12, 2022Open, Classified
Due to un-commanded movement of mobile x-ray unit.
Z-0182-2023 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Unintended System Motion4
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Bruise/Contusion1
- Unspecified Tissue Injury1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 15, 2022Unintended System Motion; Unintended System Motion
TECHNICIAN MM , RADIOLOGY TECHNOLOGIST, NIGHT SHIFT WAS COMING DOWN RAMP FROM HPMC ED TO RADIOLOGY DEPARTMENT WITH CANON SOLTUS 450ST MOBILE X-RAY UNIT "K", SN# (B)(4) AND IT SUDDENLY WENT INTO REVERSE WITHOUT WARNING. MARLON WAS PINNED AGAINST RAILING WALL AND INJURED HAND.. THE X-RAY UNIT WAS DAMAGED - THE PLASTIC FRAME UNDER HANDLE IN FRONT BROKE OFF., RADIOLOGY TECHNOLOGIST, NIGHT SHIFT WAS COMING DOWN RAMP FROM HPMC ED TO RADIOLOGY DEPARTMENT WITH CANON SOLTUS 450ST MOBILE X-RAY UNIT "K", SN# (B)(4) AND IT SUDDENLY WENT INTO REVERSE WITHOUT WARNING. MM WAS PINNED AGAINST RAILING WALL AND INJURED HAND.. THE X-RAY UNIT WAS DAMAGED - THE PLASTIC FRAME UNDER HANDLE IN FRONT BROKE OFF.
Patient problems: Bruise/Contusion
MAUDE report 15986297 ↗MalfunctionJul 22, 2022Unintended System Motion; Unintended System Motion
AFTER USING THE MOBILE, THE MOBILE LIGHT WENT RED AS THE X-RAY TECHNOLOGIST WAS BACKING OUT OF A PATIENT'S ROOM. THE X-RAY TECHNOLOGIST RE-DOCKED THE TUBE IN AN ATTEMPT TO RESET THE MOBILE MACHINE, AND HEADED BACK TO THE DEPARTMENT WITH THE MOBILE DRIVING IN "TURTLE" MODE AND THE TUBE HEAD LIGHT SHOWING RED. WHILE RETURNING BACK TO THE DEPARTMENT, MID TURN THE MOBILE WENT FULL SPEED UNCONTROLLABLY IN REVERSE AND SLAMMED INTO THE WALL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15085434 ↗MalfunctionJul 22, 2022Unintended System Motion; Unintended System Motion
ON (B)(6)2022 AT APPROXIMATELY 10:50 AM THE TECH WAS FINISHING TAKING X-RAYS IN A PATIENT ROOM ON THE 4TH FLOOR. WHILE SHE WAS ATTEMPTING TO BACK THE PORTABLE OUT OF THE ROOM SHE BEGAN HAVING DIFFICULTY WITH THE PORTABLE JERKING. SHE NOTICED A FLASHING RED LIGHT AT THE TIME. ONCE SHE GOT THE PORTABLE OUT OF THE ROOM THE TUBE HEAD LIGHT WAS STILL FLASHING RED. SHE CYCLED POWER TO THE PORTABLE AND SHE WAS STILL GETTING THE FLASHING RED LIGHT. WHEN ATTEMPTING TO TAKE THE PORTABLE BACK TO THE IMAGING DEPARTMENT THE SOLTUS MOTORS WERE NOT ENGAGING AND MOVING FORWARD. THE TECH HAD TO USE HER STRENGTH TO PUSH THE PORTABLE DOWN A HALL 20 FEET BEFORE MAKING A RIGHT TURN AND CONTINUING ABOUT ANOTHER 15 FEET BEFORE THE PORTABLE STARTED BACKING UP ON ITS OWN. THE TECH SAID THE SPEED AT WHICH THE PORTABLE WAS GOING BACKWARDS INCREASED AT WHICH TIME SHE LET GO OF THE HANDLE AND THE PORTABLE BEGAN SPINNING WHILE GOING IN REVERSE. ITS MOTION WAS ONLY STOPPED WHEN HITTING A CONCRETE PILLAR IN THE HOSPITAL HALLWAY. THE BIOMED WAS THEN CALLED. WHEN HE ARRIVED HE ACTIVATED THE E-STOP AND REMOVED THE WHEEL BOLTS THAT CONNECT THE WHEELS TO THE MOTORS. HE THEN PUSHED THE PORTABLE OUT OF THE HALLWAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15085416 ↗MalfunctionJul 22, 2022Unintended System Motion; Unintended System Motion
THE UNIT FAILED TO MOVE FORWARD WITH A RED LED ON THE TUBE HEAD INDICATING AN ERROR, THE OPERATOR THEN SHUTDOWN AND REBOOTED THE SYSTEM BUT STILL ENCOUNTERED THE SAME ISSUE. TWO TECHS ATTEMPTED TO MOVE THE PORTABLE TO GET IT OUT OF THE HALLWAY. THE PORTABLE ALL OF A SUDDEN ENGAGED BUT STARTED MOVING BACKWARD AT A HIGH RATE OF SPEED. THE TECHS LET GO OF THE PORTABLE AND IT KEPT MOVING AND RAN OVER THE TECHNOLOGIST'S FOOT. THE FACILITY INDICATED THAT THERE WAS NO SERIOUS INJURY AND MADE THE DETERMINATION THAT A VOLUNTARY 3500 FORM WAS NOT REQUIRED.
Patient problems: Unspecified Tissue Injury
MAUDE report 15085406 ↗
Other System, X-Ray, Mobile devices
- Philips MobileDiagnost wDR 2.2Sedecal · System, X-Ray, Mobile3 recalls7 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile2 recalls12 reports
- Optima XR200amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls1 report
- Optima XR220amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
- Optima XR240amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile1 recall16 reports
- MobileDaRt Evolution 503-79999-01Shimadzu Corporation · System, X-Ray, Mobile1 recall0 reports
- Optima XR240amx HD UpgradeGE Medical Systems, L.L.C. · System, x-ray, mobile1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page