Skip to content

NA 1.5T 4CH SMALL FLEX COILRecalls & reported problems

Ge Healthcare(Tianjin) Company Limited · 1.5T 4CH SMALL FLEX COIL · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 31, 2025Class IIOpen, Classified

    For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

    Z-1239-2025 · Root cause: Software design · Recall details · FDA record ↗

Other Coil, magnetic resonance, specialty devices

All Coil, magnetic resonance, specialty devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page