Skip to content

Discovery MI Digital Ready (UDI 00840682146166)Recalls & reported problems

GE Medical Systems, L.L.C. · DMI-DR PET Assy,Welded Gantry, With Labels · Class 2

All models of Discovery (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K133657); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jun 2, 2022Class IIOpen, Classified

    As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

    Z-1511-2022 · Root cause: Error in labeling · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K133657 on Mar 21, 2014 (decision: Substantially Equivalent). Submitted by Ge Medical Systems, LLC as “Discovery Pet/Ct 710”. The FDA review took 114 days. FDA record ↗

The FDA regulates product code KPS (system, tomography, computed, emission) under 21 CFR 892.1200 as a Class II device. 21 CFR 892.1200 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K133657. One submission often covers many models, so check the product and lot details in each record.

  • Z-1746-2019 · Feb 12, 2019 · Open, Classified
    Discovery PET/CT 710 Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission
  • Z-1750-2019 · Feb 12, 2019 · Open, Classified
    Discovery Ml DR Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission acti
  • Z-2579-2018 · May 24, 2018 · Terminated
    Discovery MI Digital Ready
  • Z-2191-2017 · May 19, 2017 · Terminated
    GE Healthcare Discovery Ml DR X-ray system

Common questions about the Discovery MI Digital Ready

Has the Discovery MI Digital Ready been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1511-2022 (Class II), began on Jun 2, 2022: As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Discovery MI Digital Ready FDA approved?

It was cleared through 510(k) K133657 on Mar 21, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Discovery MI Digital Ready, and is it still sold?

The FDA's device database lists GE Medical Systems, L.L.C. as the company responsible for it (brand Discovery, model MI Digital Ready). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System, tomography, computed, emission devices

All System, tomography, computed, emission devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page