Skip to content

Amendia 10-03-6550-2Recalls & reported problems

Spinal Elements · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840916102173), received between Jun 18, 2018 and Jun 18, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture1
  • Mechanical Problem1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2018Fracture; Mechanical Problem; Fracture; Mechanical Problem

    ON (B)(6) 2016, PATIENT UNDERWENT SINGLE LEVEL SPINAL FUSION SURGERY TO FUSE L4-L5. PATIENT RECEIVED AN INTERBODY DEVICE AND BILATERAL POSTERIOR FIXATION. ON (B)(6) 2017 RADIOGRAPH NOTED THAT THE PATIENT HAD POSSIBLE OSTEOLYISIS NEAR THE TIPS OF THE PEDICLE SCREWS AT L4. THE L4 VERTEBRAL BODY HAD GRADE ONE SPONDYLOLISTHESIS AND THE L5 PEDICLE SCREWS HAD FRACTURED APPROXIMATELY AT THE SECOND THREAD BELOW THE SHANK HEAD. REVISION SURGERY WAS PERFORMED TO REPLACE THE SCREWS.

    Patient problems: Failure of Implant

    MAUDE report 7610113 ↗

Other Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease devices