Skip to content

RESPONSE 5.5/6.0 SYSTEM 00-1600-0040Recalls & reported problems

Orthopediatrics Corp. · S-ROD, Ø 5.5MM COCR, 500MM, 50MM STRAIGHT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841132145890), received between Jun 22, 2025 and Jun 22, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K150600, K181390, K193100); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 22, 2025Fracture; Fracture

    HL ROD BROKE INTRA-OPERATIVELY ON TWO SEPARATE OCCASIONS WHILE SURGEON WAS USING CORONAL BENDERS TO ACHIEVE CORRECTION. IT BROKE ONCE AT BOTTOM OF THE CONSTRUCT AND AGAIN AT THE MIDDLE OF THE CONSTRUCT. IT WAS THEN REPLACED WITH S-ROD AND THE SAME THING HAPPENED- IT BROKE WHILE HE WAS CORONAL BENDING NEAR THE BOTTOM OF THE CONSTRUCT. AN OPEN/CLOSED WEDDING BAND WAS THEN USED TO ATTACH BROKEN PIECE OF ROD IN ORDER TO AVOID REPLACING FULL ROD AGAIN.

    Patient problems: Failure of Implant

    MAUDE report 22304831 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150600, K181390, K193100. One submission often covers many models, so check the product and lot details in each record.

  • Z-2742-2020 · Jul 6, 2020 · Open, Classified
    Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in sk
  • Z-0597-2018 · Aug 3, 2017 · Terminated
    Response 5.5/6.0 Spine System 7.0mm x 40mm Pedicle Screw Product Usage: Posterior non-cervical pedicle screw fixation
  • Z-0622-2016 · Nov 10, 2015 · Terminated
    UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine i

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates