Skip to content

PediNail Intramedullary Platform 00-1501-0722Recalls & reported problems

Orthopediatrics Corp. · 7MM X 220MM CHILD NAIL, LEFT · Class 2 · Implant

All models of PediNail Intramedullary Platform (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841132146101), received between Oct 2, 2024 and Oct 2, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K172583); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2024Fracture; Fracture

    NAIL WAS IMPLANTED IN SPAIN BY DOCTOR MORALEDA. BREAKAGE OF THE NAIL WAS DISCOVERED BY DR (B)(4) IN DUBAI AND PATIENT IS SCHEDULED FOR REMOVAL ON (B)(6) 2024.

    Patient problems: Failure of Implant

    MAUDE report 20356099 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172583. One submission often covers many models, so check the product and lot details in each record.

  • Z-0218-2020 · Aug 19, 2019 · Terminated
    Model No. 01-1501-0032 (Child Nail Attachment Bolt). Component of the Pediatric Nailing Platform. Product Usage: an orthopedic implant system commonly referred to as an intramedull
  • Z-0125-2020 · Aug 19, 2019 · Terminated
    OrthoPediatrics Adolescent attachment bolt used with the Pediatric Nailing Platform (PNP) targeting arm- intended as an orthopedic implant system (bolt is used to attach the tar
  • Z-0219-2020 · Aug 19, 2019 · Terminated
    Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. Product Usage: an orthopedic implant system commonly referred to as an intra

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page