Response 5.5/6.0 Spine System 01-1300-6000Recalls & reported problems
Orthopediatrics Corp. · 5.5/6.0 Rod Reducer Assembly · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00841132152027), received between Feb 17, 2020 and Mar 20, 2020 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture4
Patient problems
- No Known Impact Or Consequence To Patient3
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2020Fracture; Fracture
IT HAS BEEN REPORTED THAT DURING THE INSPECTION OF A ROD REDUCER, IT WAS FOUND TO BE FRACTURED. THE LOCATION OF THE FRAGMENTED COMPONENT IS UNKNOWN. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 9862384 ↗MalfunctionMar 4, 2020Fracture; Fracture
IT HAS BEEN REPORTED THAT DURING THE INSPECTION OF A ROD REDUCER, IT WAS FOUND TO BE FRACTURED. THE LOCATION OF THE FRAGMENTED COMPONENT IS UNKNOWN. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9788358 ↗MalfunctionMar 4, 2020Fracture; Fracture
IT HAS BEEN REPORTED THAT DURING THE INSPECTION OF A ROD REDUCER, IT WAS FOUND TO BE FRACTURED. THE LOCATION OF THE FRAGMENTED COMPONENT IS UNKNOWN. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9788351 ↗InjuryFeb 17, 2020Fracture; Fracture
IT HAS BEEN REPORTED THAT DURING CLOSURE OF A POSTERIOR SPINAL FUSION PROCEDURE, A TAB WAS IDENTIFIED TO BE MISSING FROM A ROD REDUCER. FLUOROSCOPY IDENTIFIED THAT THE TAB REMAINED IN THE PATIENT. BONE GRAFT WAS REMOVED AND THE FRACTURED PIECE WAS RECOVERED. BONE GRAFT WAS REPLACED AND INCISION CLOSED. A DELAY GREATER THAN THIRTY MINUTES WAS NOTED. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9717724 ↗
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.