Response 4.5/5.0 Spine System 00-1600-3305Recalls & reported problems
Orthopediatrics Corp. · 4.5/5.0 OPEN RIGHT 40 DEG OFFSET HOOK, 5MM · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00841132153154), received between Mar 17, 2025 and Mar 17, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Deformation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2025Material Deformation; Material Deformation
RODS HAD BEEN PLACED AND FINAL TIGHTENING WAS BEING PERFORMED WITH THE 12NM TORQUE LIMITER. WHEN THE SET SCREW OF THE HOOK WAS TIGHTENED AN AUDIBLE POP WAS HEARD. AFTER EXAMINATION IT WAS DETERMINED THE TOP 2 THREADS ON THE MEDIAL SIDE OF THE HOOK HAD SHEERED OFF. THE ROD HAD TO BE REMOVED. THE HOOK WAS REPLACED WITH A 00-1600-3315. THE NEW HOOK WAS TIGHTENED WITH A 7NM TORQUE LIMITER AND THE SET SCREW WAS FLUSH WITH THE 5.0 COCR ROD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21621900 ↗
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page