Guidewire SGW-060-01Recalls & reported problems

Galt Medical Corp. · Guidewire · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00841268100480), received between Feb 26, 2018 and Jan 4, 2019 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Entrapment of Device1
  • Material Separation1

Patient problems

  • Foreign Body In Patient2
  • Device Embedded In Tissue or Plaque1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 4, 2019Break; Device Dislodged or Dislocated

    HEALTH PROFESSIONAL SENT A COMPLAINT REPORT TO THE CUSTOMER STATING THAT "WHILE ADVANCING THE GUIDEWIRE INTO THE TIBIAL VESSEL, THE GOLD TIP BECAME DISLODGED FROM GUIDEWIRE. NO EXCESSIVE TORQUE WAS BEING APPLIED." THE CUSTOMER INFORMED MANUFACTURER THAT "THE TIP BROKE OFF DURING INSERTION AND REMAINS IN THE PATIENTS LOWER LEG (LOWER TIBIA)".

    Patient problems: Foreign Body In Patient

    MAUDE report 8218471 ↗
  • MalfunctionDec 24, 2018Break

    AS RECEIVED FROM CUSTOMER: THE 5CM GOLD TIP BROKE OFF IN THE PATIENT AS THE WIRE WAS BEING POSITIONED. WIRE WAS LEFT IN AN ACCESSORY VESSEL OF PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8194407 ↗
  • MalfunctionOct 4, 2018Break; Detachment of Device or Device Component

    AS RECEIVED FROM CUSTOMER BY END USER: 5CM GOLD TIP BROKE OFF IN THE PATIENT AS WIRE WAS TURNED TO NAVIGATE THE VESSEL, WIRE TIP BROKE OFF IN THE PATIENT WHILE POSITIONING DOWN THE LEG. APPROX. 5CM OF WIRE REMAINS IN THE PATIENT AND WAS STENTED/PINNED TO THE VESSEL WALL. ACTION TAKEN BY PHYSICIAN WAS STENTED / PINNED WIRE TIP TO VESSEL WALL AND THE ADVERSE EVENT WAS RESOLVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 7936758 ↗
  • MalfunctionFeb 26, 2018Entrapment of Device; Material Separation

    AS RECEIVED FROM END USER THROUGH THE CUSTOMER (PHYSICIAN WAS IN THE MIDDLE OF TREATING A POPLITEAL LESION, THROUGH A 5-6 SLENDER SHEATH. TIGHT AND TORTUOUS ANATOMY. HE WAS USING A CROSSER AND AS HE PULLED BACK TO THE TIBIA, HE NOTICED THE WIRE TIP HAD SEPARATED FROM THE WIRE SHAFT. (NOT SURE WHERE THE SEPARATION OCCURRED) MULTIPLE ATTEMPTS WERE MADE TO RETRIEVE THE WIRE, BUT HE ENDED UP TRAPPING THE TIP UNDER A SHEATH).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7298088 ↗

Other Wire, Guide, Catheter devices

All Wire, Guide, Catheter devices →