Skip to content

Invacare® Perfecto2™ Oxygen Concentrator IRC5PO2VRecalls & reported problems

Invacare Corporation · Class 2

All models of Invacare Perfecto2 Oxygen Concentrator (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00841447100010), received between Dec 7, 2018 and Mar 4, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Explosion” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Explosion2
  • Fire1
  • Material Deformation1
  • Material Discolored1
  • Overheating of Device1
  • Smoking1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2025Fire; Fire

    IT WAS REPORTED TO VENTEC THAT AN INVACARE PERFECTO2 OXYGEN CONCENTRATOR ALLEGEDLY CAUGHT FIRE INSIDE A PATIENT'S RESIDENCE. THE FIRE DEPARTMENT WAS DISPATCHED AND UPON ENTRY TO THE APARTMENT, OBSERVED THAT THE FIRE WAS ALREADY OUT AND THAT ONLY A LIGHT SMOKE WAS VISIBLE. THE FIRE DEPARTMENT THEN REMOVED THE OXYGEN CONCENTRATOR FROM THE APARTMENT, THROWING IT INTO THE YARD WHERE A DISTRIBUTOR WAS LATER ABLE TO RETRIEVE IT. THERE WAS PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. THERE WERE NO REPORTS OF PATIENT HARM AS A RESULT OF THE REPORTED ISSUE. THE PATIENT DECLINED MEDICAL TREATMENT AT THE SCENE. NO FURTHER DETAILS WERE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21521150 ↗
  • MalfunctionMar 18, 2019Material Discolored; Material Deformation; Explosion; Material Discolored; Material Deformation; Explosion

    THE REPORTER STATED THE TOP OF THE PRODUCT TANK AND THE PRESSURE REGULATOR HAD COME OFF IN PIECES CAUSING THE CABINET CASING TO BREAK INTO TWO PIECES, AND SIEVE DUST WAS THROUGHOUT THE UNIT. THE END USER WAS NOT IN THE ROOM AT THE TIME, AND NO INJURY WAS ALLEGED. THE UNIT WAS A RENTAL UNIT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8429229 ↗
  • MalfunctionJan 24, 2019Overheating of Device; Explosion; Overheating of Device; Explosion

    THE PROVIDER REPORTED THAT THE IRC5PO2V CONCENTRATOR MADE A LOUD NOISE, AND IT WAS DISCOVERED THAT THE PRODUCT TANK SEAL HAD BURST AND PUNCTURED THROUGH THE TOP SHROUD, WHICH DAMAGED THE HOUR METER AND SOUND BOX. HE ALLEGED THAT THE METAL CASING OF THE PRODUCT TANK CAP SHOT THROUGH THE TOP SHROUD, SENDING SHARP METAL INTO THE ROOM. THE PROVIDER ADVISED THAT THE UNIT IS A RENTAL AND WAS IN A PATIENT'S HOME WHEN THE INCIDENT OCCURRED, BUT NO INJURIES OR PROPERTY DAMAGE RESULTED FROM THE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8273945 ↗
  • MalfunctionDec 7, 2018Smoking; Smoking

    A REPAIR CENTER TECHNICIAN REPORTED THAT THE PROVIDER SENT THE IRC5PO2V CONCENTRATOR TO THEM, STATING THAT THERE WAS SMOKE COMING FROM THE UNIT. THE REPAIR TECHNICIAN ADVISED THAT WHEN HE TOOK OFF THE CABINET FILTER DOOR, THERE WAS A HOLE IN THE PLASTIC THAT WAS MELTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8141653 ↗

Other Generator, Oxygen, Portable devices

All Generator, Oxygen, Portable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page