Align ALIGN Radial Head & Lock Screw, 24mmRecalls & reported problems
Skeletal Dynamics L.L.C. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00841506100043), received between Mar 18, 2021 and Nov 24, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (9 reports).
1 other recall cover devices cleared under the same FDA submission (K092721); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component9
- Naturally Worn2
Patient problems
- Failure of Implant8
- Metal Related Pathology3
- Implant Pain2
- Foreign Body In Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 24, 2025Detachment of Device or Device Component; Naturally Worn; Detachment of Device or Device Component; Naturally Worn
PER PREVIOUS MAUDE ADVERSE EVENT REPORT (REPORT NUMBER (B)(4)), AN ALIGN RADIAL HEAD WHICH HAD BEEN IMPLANTED FOR FOUR YEARS WAS REMOVED DUE TO BEING LOOSE WITH A PARTIALLY BACKED OUT SET SCREW, REQUIRING REVISION SURGERY.
Patient problems: Failure of Implant; Metal Related Pathology
MAUDE report 23631054 ↗MalfunctionSep 13, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE SET SCREW BACKED OUT OF AN IMPLANTED ALIGN RADIAL HEAD, REQUIRING REVISION SURGERY.
Patient problems: Failure of Implant
MAUDE report 20217939 ↗MalfunctionApr 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE SET SCREW BACKED OUT OF AN IMPLANTED ALIGN RADIAL HEAD, REQUIRING REVISION SURGERY.
Patient problems: Failure of Implant
MAUDE report 19093366 ↗MalfunctionJan 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AN ALIGN RADIAL HEAD DISASSOCIATED FROM THE RADIAL STEM, REQUIRING REMOVAL SIX MONTHS AFTER INITIAL IMPLANTATION.
Patient problems: Failure of Implant; Implant Pain; Metal Related Pathology
MAUDE report 18591986 ↗MalfunctionDec 8, 2023Detachment of Device or Device Component; Naturally Worn; Detachment of Device or Device Component; Naturally Worn
AN ALIGN RADIAL HEAD WHICH HAD BEEN IMPLANTED FOR OVER FIVE YEARS WAS REMOVED DUE TO MOVEMENT OF THE RADIAL HEAD AROUND THE STEM.
Patient problems: Failure of Implant; Metal Related Pathology
MAUDE report 18297499 ↗MalfunctionNov 17, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
AN IMPLANTED RADIAL HEAD DISASSOCIATED FROM THE RADIAL STEM.
Patient problems: Failure of Implant
MAUDE report 18164916 ↗MalfunctionMay 8, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
RADIAL HEAD LOCKING SCREW LOOSENED. THE LOCKING SCREW WAS REPLACED SUCCESSFULLY, AND THERE ARE NO REPORTS OF FURTHER COMPLICATIONS.
Patient problems: Failure of Implant
MAUDE report 16892650 ↗MalfunctionApr 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
LOCKING SCREW OF RADIAL HEAD IMPLANT LOOSENED. THE AFFECTED DEVICE WAS REMOVED AND A REVISION PROCEDURE WAS CONDUCTED SUCCESSFULLY.
Patient problems: Failure of Implant
MAUDE report 16797424 ↗MalfunctionMar 18, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY 2 YEARS POST IMPLANTATION DUE TO LOOSENING OF THE PRODUCT. THE SCREW HAD BACKED OUT FROM THE RADIAL HEAD CAUSING IT TO BECOME LOOSE. THE HEAD AND SCREW WERE REPLACED ONTO THE EXISTING STEM. ATTEMPTS HAVE BEEN MADE FOR ADDITIONAL INFORMATION AND/OR THE EXPLANTED PRODUCT TO BE RETURNED BUT IS UNAVAILABLE AT THIS TIME OF FILING THE REPORT.
Patient problems: Foreign Body In Patient; Implant Pain
MAUDE report 11517590 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092721. One submission often covers many models, so check the product and lot details in each record.
- Z-0582-2015 · Oct 16, 2014 · TerminatedAlign Radial Stems, Prosthesis, Elbow, Hemi Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the
Other PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER devices
- EVOLVE 496S275Wright Medical Technology, Inc. · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall6 reports
- NA 09.405.966SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall2 reports
- NA 03.405.000SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall2 reports
- EXPLOR 11-210061Biomet Orthopedics, LLC · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.250SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.263SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.583SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
- NA 09.405.570SSynthes GmbH · PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page