Skip to content

NA 03.405.000SRecalls & reported problems

Synthes GmbH · RADIAL HEAD REPLACEMENT INSTRUMENT KIT, STERILE · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982300829), received between Jan 14, 2021 and Oct 27, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Apr 7, 2022Class IIOpen, Classified

    DePuy Synthes Radial Head Replacement System Contraindications added to Electronic Instructions for Use (e-IFU): Absolute contraindications include: Infection, sepsis, osteomyelitis.

    Z-1072-2022 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2022Break; Break

    IT WAS REPORTED ON AN UNKNOWN DATE, DURING POST OP INSPECTION, IT WAS FOUND THAT: 1) DHS COUPLING SCREW WAS BENT FROM USAGE. 2) PLASTIC HANDLE BROKE OFF AS WELL DURING RADIAL HEAD PLANNING. 3) CUTTERS WERE NOT ABLE TO CUT. 4) DRILL GUIDES¿ THREADS WERE STRIPPED. NO PATIENT CONSEQUENCE WAS REPORTED. THIS REPORT INVOLVES ONE RADIAL HEAD REPLACEMENT INSTRUMENT KIT, STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15679224 ↗
  • MalfunctionJan 14, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2020 WHILE DOING A RADIAL HEAD REPLACEMENT, AFTER USING THE PALNER, IT WAS DISCOVERED THAT SMALL PIECES OF WHITE PLASTIC HAD COME OFF DURING THE PLANNING PROCESS. ALL OF THE NOTICEABLE PIECES WERE REMOVED AND THE AREA WAS IRRIGATED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A FIVE (5) MINUTE OF SURGICAL DELAY. THE PATIENT OUTCOME IS UNKNOWN. THIS REPORT IS FOR ONE (1) RADIAL HEAD REPLACEMENT INSTRUMENT KIT, STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11170302 ↗

FDA clearance and classification

Cleared through 510(k) K183618 on May 17, 2019 (decision: Substantially Equivalent). Submitted by Ignite Orthopedics, Inc. as “Revolution Radial Head”. The FDA review took 142 days. FDA record ↗

The FDA regulates product code KWI (prosthesis, elbow, hemi-, radial, polymer) under 21 CFR 888.3170 as a Class II device. 21 CFR 888.3170 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 03.405.000S

Has the NA 03.405.000S been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1072-2022 (Class II), began on Apr 7, 2022: DePuy Synthes Radial Head Replacement System Contraindications added to Electronic Instructions for Use (e-IFU): Absolute contraindications include: Infection, sepsis, osteomyelitis. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the NA 03.405.000S?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jan 14, 2021 and Oct 27, 2022. In those reports, the most common specific device problem is “Break” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 03.405.000S?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 03.405.000S FDA approved?

It was cleared through 510(k) K183618 on May 17, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NA 03.405.000S, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 03.405.000S). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER devices

All PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page