Skip to content

Visian ICL (Implantable Collamer Lens) MICL 12.1 (UDI 00841542102742)Recalls & reported problems

Staar Surgical Company · Diopter: -9.5; Overall Diameter (Length): 12.1 mm · Class 3 · Implant

All models of Visian ICL (Implantable Collamer Lens) (96)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00841542102742), received between Sep 1, 2017 and Apr 22, 2022 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (2 reports).

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size2
  • Device Operates Differently Than Expected1
  • Misfocusing1
  • Optical Problem1

Patient problems

  • Blurred Vision2
  • Cataract1
  • Visual Disturbances1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THAT A 12.1MM, MICL12.1, -9.50 DIOPTER, IMPLANTABLE COLLAMER LENS WAS IMPLANTED INTO THE PATIENTS LEFT EYE (OS) ON (B)(6) 2021. LOW VAULT AND ANTERIOR CAPSULE CATARACT FORMATION WAS OBSERVED, LENS REMAINS IMPLANTED. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED

    Patient problems: Cataract

    MAUDE report 14190575 ↗
  • MalfunctionOct 19, 2018Misfocusing; Inadequacy of Device Shape and/or Size; Optical Problem; Misfocusing; Inadequacy of Device Shape and/or Size; Optical Problem

    THE REPORTER INDICATED LENS MICL12.1, -9.5 DIOPTER, IMPLANTED IN PATIENTS RIGHT EYE ON (B)(6) 2018 IMPROVED VAULT BUT VAULT CONTINUES TO BE HIGH. AT 3 MONTH EXAM, PATIENT WAS 20/20 UDVA, WITH A +0.25 SPH (OD), AND A 2+ VAULT. PATIENT COMPLAIN IS WITH INTERMEDIATE AND NEAR VISION AFTER LONG DAYS. PATIENT HAS CHOSEN NOT TO EXPLANT LENS. ADDITIONAL INFORMATION, PILO 1% AND LIMIFY HAVE BENN OF MINIMAL HELP.

    Patient problems: Blurred Vision; Visual Disturbances

    MAUDE report 7987063 ↗
  • MalfunctionSep 1, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED AN MICL12.1, -9.5 DIOPTER, IMPLANTABLE COLLAMER LENS IN THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2017. REFRACTIVE SURPRISE WAS NOTICED AS THE PATIENT IS NOW APPROXIMATELY +1.25D SEQ POST-OP. THE SURGEON CONFIRMED THAT THE ICL IS SITTING IN THE CORRECT POSITION AND NOT INVERTED. HE ALSO VERIFIED THAT THE CYCLO WAS NOT SIGNIFICANTLY DIFFERENT THAT THE MANIFEST REFRACTION. ATTEMPTS AT OBTAINING ADDITIONAL INFORMATION HAVE BEEN UNSUCCESSFUL.

    Patient problems: Blurred Vision

    MAUDE report 6839063 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.

Other Lens, Intraocular, Phakic devices

All Lens, Intraocular, Phakic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page