Visian ICL (Implantable Collamer Lens) MICL 12.6 (UDI 00841542102919)Recalls & reported problems
Staar Surgical Company · Diopter: -4.5; Overall Diameter (Length): 12.6 mm · Class 3 · Implant
All models of Visian ICL (Implantable Collamer Lens) (96)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00841542102919), received between Nov 9, 2016 and Feb 26, 2021 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected2
- Inadequacy of Device Shape and/or Size2
- Optical Problem1
- Unintended Movement1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- Halo1
- No Clinical Signs, Symptoms or Conditions1
- Visual Disturbances1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryFeb 26, 2021Inadequacy of Device Shape and/or Size; Optical Problem; Inadequacy of Device Shape and/or Size; Optical Problem
IT WAS REPORTED THAT A 12.6MM, MICL12.6, -4.5 DIOPTER, IMPLANTABLE COLLAMER LENS LENS WAS GOING TO BE EXPLANTED DUE EXCESSIVE VAULT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.
Patient problems: Halo; No Clinical Signs, Symptoms or Conditions; Visual Disturbances
MAUDE report 11384756 ↗MalfunctionJul 22, 2019Unintended Movement; Unintended Movement
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM MICL12.6, -4.5 DIOPTER IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S EYE ON (B)(6) 2019. THE SURGEON INDICATED THAT AFTER INSERTION, THE LENS "UNFOLDED UPSIDE DOWN" IN THE EYE. INTRAOPERATIVELY, THE LENS WAS SUCCESSFULLY EXCHANGED WITH AN ALTERNATE LENS. REPORTEDLY, NO PATIENT INJURY. "PATIENT DOING WELL, WITH NO EVIDENCE OF PROBLEMS IN THE EYE."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8813784 ↗MalfunctionDec 13, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A MICL12.6, -4.5 DIOPTER, IMPLANTABLE COLLAMER LENS IN THE RIGHT EYE (OD) ON (B)(6) 2017. DURING IMPLANTATION, THE PATIENT'S EYE ROLLED. THE SURGEON EXPLANTED AND EXCHANGED THE LENS INTRAOPERATIVELY WITH A BACK-UP LENS AND IT WAS SUCCESSFUL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7114229 ↗MalfunctionMar 24, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
THE REPORTER STATED A 12.6MM MICL12.6 IMPLANTABLE COLLAMER LENS, -4.5 DIOPTER, TORE DURING THE LOADING PROCESS AND THERE WAS NO PATIENT CONTACT. THE BACKUP LENS WAS IMPLANTED. THE REPORTER STATED THE CAUSE OF THE EVENT WAS UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6432876 ↗InjuryNov 9, 2016Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE REPORTER STATED THE SURGEON INSERTED A 12.6MM MICL12.6 IMPLANTABLE COLLAMER LENS, -4.5 DIOPTER, IN THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2016. THE LENS WAS NOTED TO BE TOO BIG AND WAS REMOVED DURING THE SAME SURGERY WITH NO PATIENT INJURY OR COMPLICATIONS. THE LENS WAS NOT EXCHANGED FOR ANOTHER LENS. THE REPORTER STATED THE CAUSE OF THE EVENT WAS UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6089911 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.
- Z-1480-2026 · Jan 21, 2026 · Open, ClassifiedEVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
- Z-0469-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
- Z-0474-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
- Z-0468-2024 · Sep 29, 2023 · Open, ClassifiedEVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
- Z-0466-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
- Z-0465-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
- Z-0467-2024 · Sep 29, 2023 · Open, ClassifiedEVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
- Z-0464-2024 · Sep 29, 2023 · Open, ClassifiedEVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
Other Lens, Intraocular, Phakic devices
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall12 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-11.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-5.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-8.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall11 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-8.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 13.2 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.0)Staar Surgical Company · Lens, Intraocular, Phakic1 recall9 reports
- EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-9.5)Staar Surgical Company · Lens, Intraocular, Phakic1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗