Visian ICL (Implantable Collamer Lens) MICL 13.7 (UDI 00841542103602)Recalls & reported problems

Staar Surgical Company · Diopter: -12.0; Overall Diameter (Length): 13.7 mm · Class 3 · Implant

All models of Visian ICL (Implantable Collamer Lens) (96)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00841542103602), received between Oct 12, 2016 and Sep 7, 2021 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (4 reports).

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size4
  • Device Operates Differently Than Expected1
  • Material Opacification1

Patient problems

  • No Code Available3
  • Blurred Vision1
  • Intraocular Pressure, Delayed, Uncontrolled1
  • No Clinical Signs, Symptoms or Conditions1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 7, 2021Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.7MM MICL13.7 IMPLANTABLE COLLAMER LENS, -12.00 DIOPTER, IN THE PATIENTS EYE ON (B)(6) 2021. THE LENS WAS EXPLANTED THE SAME DAY DUE TO THE LENS WAS TOO BIG. THERE WAS NO PATIENT INJURY. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12437086 ↗
  • InjuryFeb 28, 2019Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.7MM MICL13.7 IMPLANTABLE COLLAMER LENS, -12.0 DIOPTER, IN THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2019. THE LENS WAS EXPLANTED ON (B)(6) 2019 DUE TO EXCESSIVE VAULTING AND SIGNIFICANT REDUCTION OF IRIDO-CORNEAL ANGLES. THE LENS WAS EXCHANGED FOR A SHORTER LENS AND THE PROBLEM WAS RESOLVED.

    Patient problems: No Code Available

    MAUDE report 8379574 ↗
  • MalfunctionFeb 19, 2019Material Opacification; Material Opacification

    THE REPORTER INDICATED THAT AFTER OPENING THE LENS VIAL IN ORDER TO IMPLANT A 13.7MM MICL13.7 IMPLANTABLE COLLAMER LENS, -12.0 DIOPTER, INTO THE PATIENT'S RIGHT (OD) EYE, IT WAS NOTICED THAT THE LENS WAS "CLOUDY." THIS OCCURRED ON (B)(6) 2019. THERE WAS NO PATIENT CONTACT WITH THIS LENS. THE REPORTER STATES, "ICL HAD OPACIFICATION."

    Patient problems: No Patient Involvement

    MAUDE report 8350795 ↗
  • InjuryDec 14, 2017Inadequacy of Device Shape and/or Size; Device Operates Differently Than Expected; Inadequacy of Device Shape and/or Size; Device Operates Differently Than Expected

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A MICL13.7, -12.0 DIOPTER, IMPLANTABLE COLLAMER LENS IN THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2017. ON (B)(6) 2017, THE LENS WAS EXPLANTED DUE TO DECREASE VISUAL ACUITY. AT THE TIME OF SUBMISSION, ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN UNSUCCESSFUL.

    Patient problems: Blurred Vision; No Code Available

    MAUDE report 7119003 ↗
  • InjuryOct 12, 2016Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER STATED THE SURGEON IMPLANTED A 13.7MM MICL13.7 IMPLANTABLE COLLAMER LENS, -12.0 DIOPTER, IN THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2016. THE LENS WAS EXPLANTED ON (B)(6) 2016 DUE TO VAULTING. THE LENS WAS EXCHANGED FOR A SHORTER LENS.

    Patient problems: Intraocular Pressure, Delayed, Uncontrolled; No Code Available

    MAUDE report 6024713 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.

Other Lens, Intraocular, Phakic devices

All Lens, Intraocular, Phakic devices →