Skip to content

Visian ICL (Implantable Collamer Lens) MICL 13.7 (UDI 00841542103619)Recalls & reported problems

Staar Surgical Company · Diopter: -12.5; Overall Diameter (Length): 13.7 mm · Class 3 · Implant

All models of Visian ICL (Implantable Collamer Lens) (96)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00841542103619), received between Jan 26, 2017 and Jun 4, 2019 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Inadequacy of Device Shape and/or Size” (3 reports).

8 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Inadequacy of Device Shape and/or Size3
  • Device Operates Differently Than Expected1
  • No Apparent Adverse Event1

Patient problems

  • No Code Available3
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 4, 2019Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.7MM MICL13.7 IMPLANTABLE COLLAMER LENS, -12.5 DIOPTER, IN THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2019. THE REPORTER INDICATED THE SURGEON IS PLANNING TO EXPLANT THE LENS DUE TO BEING TOO LONG. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.

    Patient problems: No Code Available; No Known Impact Or Consequence To Patient

    MAUDE report 8668925 ↗
  • InjuryFeb 28, 2019Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE REPORTER INDICATED THE SURGEON IMPLANTED A 13.7MM MICL13.7 IMPLANTABLE COLLAMER LENS, -12.5 DIOPTER, IN THE PATIENTS LEFT EYE (OS) ON (B)(6) 2019. THE LENS WAS EXPLANTED ON (B)(6) 2019 DUE TO EXCESSIVE VAULTING AND SIGNIFICANT REDUCTION OF IRIDO-CORNEAL ANGLES. THE LENS WAS EXCHANGED FOR A SHORTER LENS AND THE PROBLEM WAS RESOLVED.

    Patient problems: No Code Available

    MAUDE report 8379587 ↗
  • MalfunctionOct 2, 2018No Apparent Adverse Event; No Apparent Adverse Event

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 13.7MM MICL13.7 IMPLANTABLE COLLAMER LENS, -12.5 DIOPTER, INTO THE PATIENT'S EYE ON (B)(6) 2018. INTRAOPERATIVELY, THE LENS WAS SUCCESSFULLY EXCHANGED WITH AN ALTERNATE LENS. THE REASON FOR THE EXCHANGE IS THAT IT "WAS NOT POSITIONED CORRECTLY." "NO PATIENT INJURY" IS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7927074 ↗
  • InjurySep 12, 2018Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    THE DOCTOR INDICATED THAT HE IMPLANTED A 13.7 MM MICL13.7, -12.5 DIOPTER, IMPLANTABLE COLLAMER LENS INTO THE PATIENTS RIGHT EYE (OD) ON (B)(6) 2018. ON (B)(6) 2018 THE LENS WAS REMOVED. IT WAS REPORTED THAT THE LENS WAS EXCHANGE 3D AND THE PROBLEM WAS RESOLVED.

    Patient problems: No Code Available

    MAUDE report 7868380 ↗
  • MalfunctionJan 26, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    THE REPORTER STATED THE SURGEON INSERTED A 13.7MM MICL13.7 IMPLANTABLE COLLAMER LENS, -12.5 DIOPTER, AND THE LENS TORE. THE LENS WAS REMOVED WITHIN THE SAME SURGERY WITH NO PATIENT INJURY, NO ENLARGED INCISION OR SUTURES. THE BACKUP LENS WAS IMPLANTED WITH NO PROBLEM. THE REPORTER STATED THE CAUSE OF THE EVENT WAS UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6281925 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.
  • Z-0465-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_12.6. Phakic Intraocular lens.
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.

Other Lens, Intraocular, Phakic devices

All Lens, Intraocular, Phakic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page