EVO/EVO+VISIAN Implantable Collamer Lens VICM5 12.6 (-11.5) (REF VICM5 12.6)Recalls & reported problems

Staar Surgical Company · VICM5 - Diopter: -11.5; Overall Diameter (Length): 12.6 · Class 3 · Implant

All models of EVO/EVO+VISIAN Implantable Collamer Lens (1030)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00841542119214), received between Aug 5, 2022 and Sep 12, 2023 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

7 other recalls cover devices cleared under the same FDA submission (P030016); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Sep 29, 2023Class IIOpen, Classified

    Their is a potential that intraocular implant devices may not meet specifications.

    Z-0465-2024 · Root cause: Software design (manufacturing process) · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Apparent Adverse Event3
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Inadequacy of Device Shape and/or Size1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 12, 2023Inadequacy of Device Shape and/or Size; Device Dislodged or Dislocated; Inadequacy of Device Shape and/or Size; Device Dislodged or Dislocated

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM5_12.6, -11.5 DIOPTER IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2023. THE PATIENT EXPERIENCED LOW VAULTING WITH ROTATION. ON (B)(6) 2023 THE LENS WAS EXCHANGED WITH A LONGER LENGTH LENS AND THE PROBLEM WAS RESOLVED. THE CAUSE OF THE EVENT WAS REPORTED AS USER ERROR WITH NO DEVICE CAUSALITY AND NOTED "LOW VAULT. I HAVE TYPICALLY IMPLANTED LOWER VAULT ICL IN THESE CASES AND HAVEN'T HAD ISSUES IN THE PAST".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17733898 ↗
  • MalfunctionJul 28, 2023No Apparent Adverse Event; No Apparent Adverse Event

    THE REPORTER INDICATED THAT 12.6MM VICM512.6 IMPLANTABLE COLLAMER LENS OF A -11.50 DIOPTER TORE/BROKE DURING INJECTION INTO THE PATIENT'S EYE. THE LENS WAS IMPLANTED AND REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17421355 ↗
  • MalfunctionMar 8, 2023No Apparent Adverse Event; No Apparent Adverse Event

    THE REPORTER INDICATED THAT ON (B)(6) 2023 A 12.6MM VICM5_12.6 -11.50 DIOPTER IMPLANTABLE COLLAMER LENS WAS NOTED TO BE DAMAGED (TORN) WHILE INJECTING INTO THE PATIENTS LEFT EYE (OS). PER REPORTER, THE DAMAGE OCCUR DURING LOADING. LENS STUCK IN CARTRIDGE/DELIVERY SYSTEM WAS ALSO REPORTED. THERE WAS PATIENT CONTACT WITH NO PATIENT INJURY. A REPLACEMENT LENS WAS IMPLANTED WITHIN THE SAME SURGERY AND THE PROBLEM RESOLVED. IN THE REPORTER'S OPINION, THE DEVICE IS THE CAUSE OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16500625 ↗
  • InjuryDec 5, 2022Unintended Movement; Unintended Movement

    THE REPORTER INDICATED THAT THE SURGEON IMPLANTED A 12.6MM VICM5_12.6; -11.50 DIOPTER IMPLANTABLE COLLAMER LENS INTO THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2022. THE LENS HAD FLIPPED DURING INSERTION. THE LENS WAS EXPLANTED AND REPLACED WITH A BACK UP LENS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE NOT BEEN SUCCESSFUL. IF ADDITIONAL INFORMATION IS RECEIVED A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15921134 ↗
  • InjuryAug 5, 2022Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event

    THE REPORTER INDICATED THE SURGEON INSERTED AND REMOVED A 12.6MM VICM5_12.6 IMPLANTABLE COLLAMER LENS, -11.50 DIOPTER, WITHIN THE SAME SURGERY DUE TO THE LENS WAS STUCK IN CARTRIDGE OR INJECTION SYSTEM. THERE WAS NO PATIENT INJURY. THE LENS WAS REPLACED WITH A MICL12.6 LENS AND THE PROBLEM WAS RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15178131 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030016. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2026 · Jan 21, 2026 · Open, Classified
    EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
  • Z-0467-2024 · Sep 29, 2023 · Open, Classified
    EVO +VISIAN Implantable Collamer Lens, REF: VICM5_13.7.Phakic Intraocular lens.
  • Z-0474-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.6. Phakic Intraocular lens.
  • Z-0464-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF:VICM5_12.1. Phakic Intraocular lens.
  • Z-0469-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens REF: VICMO13.2. Phakic Intraocular lens.
  • Z-0466-2024 · Sep 29, 2023 · Open, Classified
    EVO+VISIAN Implantable Collamer Lens, REF: VICM5_13.2. Phakic Intraocular lens.
  • Z-0468-2024 · Sep 29, 2023 · Open, Classified
    EVO VISIAN Implantable Collamer Lens, REF: VICMO12.1. Phakic Intraocular lens.

Other Lens, Intraocular, Phakic devices

All Lens, Intraocular, Phakic devices →