Skip to content

Mammotome elite MEP10Recalls & reported problems

Devicor Medical Products, Inc. · Probe - CE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00841911100881), received between May 22, 2019 and Aug 24, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Failure to Obtain Sample” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Obtain Sample1
  • Unsealed Device Packaging1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, THE EVENT OCCURRED DURING BREAST BIOPSY PROCEDURE. AFTER THE TISSUES WERE COLLECTED, THE SAMPLE CUP WAS OPENED AND A PIECE OF PLASTIC WAS FOUND, WHICH WAS NOT NOTICED DURING PRIMING. THE SIZE OF THE FOREIGN OBJECT WAS ABOUT 5 MILLIMETERS. THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT COMPLICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17613006 ↗
  • MalfunctionAug 1, 2019Failure to Obtain Sample; Failure to Obtain Sample

    DEVICOR MEDICAL PRODUCTS, INC. RECEIVED A REPORT FROM AFFILIATE, DEVICOR MEDICAL (B)(4)" IN THE INITIALIZING PROCESS, THE PROBE WAS ATTACHED TO THE HOLSTER PROPERLY AND COMPLETED PRIMING, HOWEVER THE SPECIMEN WAS NOT TRANSPORTED TO THE CHAMBER". THIS HAS BEEN DOCUMENTED IN OUR SYSTEM AS RECORD # (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8852881 ↗
  • MalfunctionMay 22, 2019Unsealed Device Packaging; Unsealed Device Packaging

    DEVICOR MEDICAL PRODUCTS, INC. RECEIVED A REPORT FROM OUR AFFILIATE (B)(4)., STATING, PACKAGE DAMAGED RESULTING IN BREACH OF STERILE SEAL. NO PATIENT INVOLVEMENT. THIS HAS BEEN DOCUMENTED IN OUR SYSTEM AS RECORD # (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8631240 ↗

Other Instrument, Biopsy devices

All Instrument, Biopsy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page