Merge Unity PACS 11.2Recalls & reported problems
Merge Healthcare Solutions Inc. · Standalone software application · Class 2
All models of Merge Unity PACS (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00842000100331), received between Jan 1, 2019 and Oct 27, 2020 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Loss of Data” (4 reports).
6 other recalls cover devices cleared under the same FDA submission (K041935); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Loss of Data4
- Computer Software Problem1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 27, 2020Loss of Data; Loss of Data
MERGE UNITY PACS IS A MEDICAL IMAGE AND INFORMATION MANAGEMENT SYSTEM THAT IS USED FOR VIEWING, SELECTION, PROCESSING, PRINTING, TELECOMMUNICATIONS, AND MEDIA INTERCHANGE OF MEDICAL IMAGES FROM A VARIETY OF DIAGNOSTIC IMAGING SYSTEMS. ON (B)(6) 2020, A CUSTOMER CONTACTED MERGE HEALTHCARE TO INFORM US THAT THEY WERE UNABLE TO VIEW AN EXAM REPORT FROM A RECENT EXAM. UPON INVESTIGATION AND TROUBLESHOOTING BY MERGE HEALTHCARE SUPPORT, IT WAS DETERMINED THAT THE REPORT DID NOT SAVE DUE TO A WORKSTATION CRASH. A BLANK REPORT DOCUMENT WAS MANUALLY PLACED IN THE FOLDER AND THE REPORT WAS RE-DICTATED AND SUBSEQUENTLY VIEWABLE AS EXPECTED. THIS HAS THE POTENTIAL TO DELAY PATIENT TREATMENT AND/OR DIAGNOSIS. THERE HAVE BEEN NO REPORTS OF PATIENT INJURY OR HARM AS A RESULT OF THIS ISSUE. REFERENCE (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10744059 ↗MalfunctionSep 4, 2020Computer Software Problem; Loss of Data; Computer Software Problem; Loss of Data
MERGE UNITY PACS IS A MEDICAL IMAGE AND INFORMATION MANAGEMENT SYSTEM THAT IS USED FOR VIEWING, SELECTION, PROCESSING, PRINTING, TELECOMMUNICATIONS, AND MEDIA INTERCHANGE OF MEDICAL IMAGES FROM A VARIETY OF DIAGNOSTIC IMAGING SYSTEMS. ON (B)(6) 2020, A CUSTOMER CONTACTED MERGE HEALTHCARE TO INFORM US THAT THEY WERE UNABLE TO VIEW AN EXAM REPORT FROM A RECENT EXAM. UPON INVESTIGATION AND TROUBLESHOOTING BY MERGE HEALTHCARE SUPPORT, IT WAS DETERMINED VIA APPLICATION LOGS THAT THE REPORT NEVER UPLOADED BECAUSE THE C DRIVE ON THE LOCAL STATION WAS FULL, WHICH PREVENTED THE REPORT FROM SAVING. SUPPORT ASSISTED THE CUSTOMER IN RE-GENERATING THE REPORT AND RECOMMENDED REVIEWING THE C DRIVE OF THE LOCAL STATION. THIS HAS THE POTENTIAL TO DELAY PATIENT TREATMENT AND/OR DIAGNOSIS. THERE HAVE BEEN NO REPORTS OF PATIENT INJURY OR HARM AS A RESULT OF THIS ISSUE. REFERENCE COMPLAINT (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10496088 ↗MalfunctionMar 19, 2019Loss of Data; Loss of Data
MERGE UNITY PACS IS A MEDICAL IMAGE AND INFORMATION MANAGEMENT SYSTEM THAT IS USED FOR VIEWING, SELECTION, PROCESSING, PRINTING, TELECOMMUNICATIONS, AND MEDIA INTERCHANGE OF MEDICAL IMAGES FROM A VARIETY OF DIAGNOSTIC IMAGING SYSTEMS. ON (B)(6) 2019 THE CUSTOMER REPORTED AN EXAM WAS MARKED READ, HOWEVER, THERE IS NO EXAM REPORT. THE REPORT WAS SAVED, BUT NOT ASSOCIATED PROPERLY WITH THE EXAM. A CAUSE OF THE FAILED ASSOCIATION WAS NOT ABLE TO BE DETERMINED. AN INVESTIGATION SHOWED THE REPORT DID NOT UPLOAD SUCCESSFULLY AND HAD TO BE READ AGAIN BY THE DOCTOR. THERE WAS NO REPORTED ADVERSE EVENT TO THE PATIENT. HOWEVER, WHILE IMAGES ARE AVAILABLE, A FINALIZED REPORT NOT BEING AVAILABLE FOR SUBSEQUENT REVIEW BY THE PATIENT'S PHYSICIAN COULD POTENTIALLY RESULT IN A DELAY IN CARE THAT COULD LEAD TO HARM. REFERENCE SALES FORCE (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8431941 ↗MalfunctionJan 1, 2019Loss of Data; Loss of Data
MERGE UNITY PACS IS A MEDICAL IMAGE AND INFORMATION MANAGEMENT SYSTEM THAT IS USED FOR VIEWING, SELECTION, PROCESSING, PRINTING, TELECOMMUNICATIONS, AND MEDIA INTERCHANGE OF MEDICAL IMAGES FROM A VARIETY OF DIAGNOSTIC IMAGING SYSTEMS. ON DECEMBER 13, 2018 THE CUSTOMER REPORTED AN EXAM WAS COMPLETED, HOWEVER, THERE IS NO EXAM REPORT. AN INVESTIGATION SHOWED THE REPORT DID NOT UPLOAD SUCCESSFULLY AND HAD TO BE READ AGAIN BY THE DOCTOR. THE EXAM WAS READ, THERE WAS AN UPLOAD ERROR AT THE TIME OF THE SAVE EVENT. THERE WAS NO REPORTED ADVERSE EVENT TO A PATIENT. A MISSING REPORT DOES HAVE THE POTENTIAL TO DELAY PATIENT CARE. (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8207334 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K041935. One submission often covers many models, so check the product and lot details in each record.
- Z-3105-2017 · Aug 14, 2017 · TerminatedMerge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and informatio
- Z-2566-2018 · Aug 7, 2017 · TerminatedMerge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is Merge Healthcare, Hartland, WI. Merge Unity is a medical image and informati
- Z-1426-2017 · Apr 8, 2016 · TerminatedDR Systems Unity PACS software, now known as Merge Unity PACS
- Z-1385-2017 · Apr 8, 2016 · TerminatedDR Systems Unity PACS software, now known as Merge Unity PACS. Product Usage: Merge Unity PACS is a medical image and information management system that allows viewing, selection,
- Z-0939-2017 · Oct 15, 2015 · TerminatedDR Systems Unity PACS software, now known as Merge Unity PACS software.
- Z-2551-2016 · Jul 23, 2015 · TerminatedMerge Unity Z3D software. Merge Unity PACS (formerly DR Systems PACS) is a medical image and information management system that allows viewing, selection, processing, printing, te
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page