CADstream 6.5Recalls & reported problems
Merge Healthcare Solutions Inc. · Standalone software application · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00842000101116), received between Jan 18, 2025 and Jan 18, 2025 — classified as 11 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (11 reports).
3 other recalls cover devices cleared under the same FDA submission (K092954); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method11
- Use of Incorrect Control/Treatment Settings11
Patient problems
- Breast Mass11
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available10
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE-MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED DIAG
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187617 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED DIAG
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187616 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED DIAG
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187615 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED DIAG
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187613 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED DIAG
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187612 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED DIAG
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187611 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED DIAG
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187610 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, (B)(6) HEALTHCARE RECEIVED A COMPLAINT FROM A (B)(6) CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. (B)(6)HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. (B)(6) TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, (B)(6) HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187609 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON 12/06/2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S DIAGNOSTIC ASSESSMENT CATEGORY WAS REVISED. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON THE REVISED DIAG
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187608 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON (B)(6)2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE-MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE RE-PROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. THERE HAS BEEN NO REPORT OF PATIENT HARM.
Patient problems: Breast Mass; No Clinical Signs, Symptoms or Conditions
MAUDE report 21187607 ↗InjuryJan 18, 2025Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method; Use of Incorrect Control/Treatment Settings; Improper or Incorrect Procedure or Method
ON (B)(6)2024, MERGE HEALTHCARE RECEIVED A COMPLAINT FROM A MERGE CADSTREAM CUSTOMER ALLEGING THAT THE SYSTEM DID NOT HIGHLIGHT IMAGES FROM A PATIENT WITH MALIGNANCY. MERGE HEALTHCARE TECHNICAL SUPPORT CONNECTED TO THE CUSTOMER'S SYSTEM TO TROUBLESHOOT THE ISSUE. TECHNICAL SUPPORT FOUND THAT THE CUSTOMER'S DEFAULT STUDY PREFERENCE (DSP) HAD NOT BEEN OPTIMIZED FOR IMAGE ANALYSIS WITH A PRE-MASK SEQUENCE AND THE DYNAMIC SEQUENCE WITH PEAK CONTRAST. MERGE TECHNICAL SUPPORT WORKED WITH THE CUSTOMER AND ADJUSTED THE DSP CONFIGURATION SETTINGS TO ENHANCE IMAGE ANALYSIS. THE PATIENT'S STUDY WAS RE-PROCESSED BY CADSTREAM WITH THE OPTIMIZED DSP CONFIGURATION. THE CUSTOMER REVIEWED THE REPROCESSED STUDY AND CONFIRMED THAT THE HIGHLIGHTING IS WORKING AS EXPECTED. ON 12/19/2024, MERGE HEALTHCARE RECEIVED ADDITIONAL INFORMATION FROM THE CUSTOMER INDICATING THAT THEY HAD RE-PROCESSED AND REVIEWED IMAGES FOR PATIENTS WHOSE IMAGES HAD BEEN REVIEWED BASED ON THE PRIOR DSP CONFIGURATION. AFTER A RADIOLOGIST RE-REVIEWED A PATIENT'S IMAGES WITH THE OPTIMIZED DSP CONFIGURATION, THE PATIENT'S TREATMENT PLAN WAS REVISED TO INCLUDE ADDITIONAL TESTING. THERE HAS BEEN NO REPORT OF PATIENT HARM BECAUSE ON TH
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Breast Mass
MAUDE report 21187606 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092954. One submission often covers many models, so check the product and lot details in each record.
- Z-0137-2017 · Jan 30, 2016 · TerminatedCADstream software. The firm name on the label is Merge Healthcare, Hartland, WI.
- Z-2871-2016 · Nov 1, 2012 · TerminatedCADstream software. Product Usage: CADstream is an image processing system designed to assist in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) st
- Z-0118-2017 · Oct 19, 2011 · TerminatedCADstream software Product Usage: CADstream is an image processing system designed to assist in the visualization, analysis, and reporting of magnetic resonance imaging (MRI) st
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page