Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842188123849), received between Dec 15, 2022 and Dec 15, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Dec 22, 2022Class IIOpen, Classified
Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.
Z-1101-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation2
Patient problems
- Damage to Ligament(s)2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 15, 2022Material Separation; Material Separation
JIM FASSETT WAS NOTIFIED ON 02DEC2022 THAT THE PULL SUTURE IN STA001K SEPARATED FROM THE MEDIAL BUTTON DURING SURGERY. IT WAS REPORTED THAT THE MEDIAL BUTTON WAS THROUGH THE FIBULAR AND TIBIAL BONE TUNNELS AND SUCCESSFULLY DEPLOYED ON THE MEDIAL CORTEX. THE SALES REP REPORTED THAT THIS DID NOT CAUSE A DELAY IN SURGERY NOR DID IT CAUSE THE PATIENT ADDITIONAL HARM. THERE ARE NO PICTURES AVAILABLE AND THE DEVICE IS NOT AVAILABLE FOR RETURN.
Patient problems: Damage to Ligament(s)
MAUDE report 15993141 ↗MalfunctionDec 15, 2022Material Separation; Material Separation
JIM FASSETT WAS NOTIFIED ON 02DEC2022 THAT THE PULL SUTURE IN STA001K SEPARATED FROM THE MEDIAL BUTTON DURING SURGERY. IT WAS REPORTED THAT THE MEDIAL BUTTON WAS THROUGH THE FIBULA BONE TUNNEL, BUT NOT YET IN THE TIBIAL BONE TUNNEL ALLOWING THE SURGEON TO RETRIEVE THE DEVICE ENOUGH TO APPLY AN OFF-THE-SHELF PASSING SUTURE / NEEDLE THAT WAS RETHREADED THROUGH THE PULL SUTURE HOLE IN THE LEADING EDGE OF THE BUTTON. THE SALES REP REPORTED THAT THIS DID NOT CAUSE A DELAY IN SURGERY NOR DID IT CAUSE THE PATIENT ADDITIONAL HARM. THERE ARE NO PICTURES AVAILABLE AND THE DEVICE IS NOT AVAILABLE FOR RETURN.
Patient problems: Damage to Ligament(s)
MAUDE report 15993140 ↗
FDA clearance and classification
The FDA regulates product code HTN (washer, bolt nut) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the RipCord STA001K
Has the RipCord STA001K been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1101-2023 (Class II), began on Dec 22, 2022: Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the RipCord STA001K?
The FDA's MAUDE database lists 2 adverse event reports for this device, received on Dec 15, 2022. In those reports, the most common specific device problem is “Material Separation” (2); the most common patient problem is “Damage to Ligament(s)” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the RipCord STA001K?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the RipCord STA001K MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the RipCord STA001K, and is it still sold?
The FDA's device database lists Trimed, INC. as the company responsible for it (brand RipCord, model STA001K). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other WASHER, BOLT NUT devices
- FIBULINK FGS-1000Synthes GmbH · Washer, Bolt Nut116 reports
- Arthrex® AR-2372BLOArthrex, Inc. · WASHER, BOLT NUT91 reports
- Arthrex® AR-8925SSArthrex, Inc. · WASHER, BOLT NUT79 reports
- Arthrex® AR-8925TArthrex, Inc. · WASHER, BOLT NUT70 reports
- Arthrex® AR-8926TArthrex, Inc. · WASHER, BOLT NUT60 reports
- Arthrex® AR-7000-18TArthrex, Inc. · WASHER, BOLT NUT1 recall0 reports
- Arthrex® AR-8926SSArthrex, Inc. · WASHER, BOLT NUT35 reports
- Arthrex® AR-8919DSArthrex, Inc. · WASHER, BOLT NUT27 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page