Skip to content

Arthrex® AR-7000-18TRecalls & reported problems

Arthrex, Inc. · SUTURE WASHER, TITANIUM · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Aug 12, 2016Class IIITerminated

    The device was assembled incorrectly. Suture and washer were not assembled; components were placed loose in the package.

    Z-2918-2016 · Root cause: Packaging process control · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code HTN (washer, bolt nut) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Arthrex® AR-7000-18T

Has the Arthrex® AR-7000-18T been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2918-2016 (Class III), began on Aug 12, 2016: The device was assembled incorrectly. Suture and washer were not assembled; components were placed loose in the package. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Arthrex® AR-7000-18T MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Arthrex® AR-7000-18T, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-7000-18T). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other WASHER, BOLT NUT devices

All WASHER, BOLT NUT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page