WEB Detachment Controller WDC-2-MVIRecalls & reported problems
Microvention, Inc. · WEB Detachment Controller · Class 3
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Sep 21, 2022Open, Classified
Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels
Z-0170-2023 · Root cause: Process control · Recall details · FDA record ↗
Other Intrasaccular Flow Disruption Device devices
- WEB W5-4-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death23 reports
- WEB 17 SL 6x2 W5-6-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death20 reports
- WEB W5-5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device20 reports
- WEB W5-4.5-2-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device19 reports
- WEB W5-7-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death16 reports
- WEB W5-6-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device1 death15 reports
- WEB W5-5-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device14 reports
- WEB Aneurysm Embolization System W2-8-3-MVI-3Microvention, Inc. · Intrasaccular Flow Disruption Device14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page