Stent FRED5514-PMARecalls & reported problems
Microvention, Inc. · FRED 27 · Class 3 · Implant
All models of Stent (63)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00842429106853), received between Apr 2, 2020 and Aug 26, 2021 — classified as 5 injuries.
The most frequently reported specific device problem is “Activation Failure” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Jun 7, 2023Open, Classified
Intracranial aneurysm flow diverter device may be manufactured with an incorrect inner stent length and may not perform as intended.
Z-2238-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure4
- Adverse Event Without Identified Device or Use Problem1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
- Stroke/CVA1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A FRED STENT WAS IMPLANTED IN A PATIENT ON (B)(6) 2020, THE PATIENT PRESENTED TO THE HOSPITAL WITH STROKE-LIKE SYMPTOMS ON (B)(6) 2021. THERE WAS A COMPLETE OCCLUSION OF THE INTERNAL CAROTID ARTERY. THE PATIENT WAS ON ASA AND PLAVIX FOR 3 MONTHS POST PROCEDURE.
Patient problems: Stroke/CVA
MAUDE report 12370807 ↗InjuryMay 1, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT A FRED STENT DID NOT FULLY OPEN AFTER BEING IMPLANTED IN THE PATIENT. A BALLOON CATHETER WAS THEN USED TO FULLY OPEN THE STENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11758658 ↗InjuryDec 16, 2020Activation Failure; Activation Failure
IT WAS REPORTED THAT AFTER DEPLOYMENT OF THE FRED, THE STENT DID NOT APPEAR TO BE COMPLETELY OPEN AT THE PROXIMAL END, WHICH LIMITED FLOW. A BALLOON INFLATION WAS PERFORMED AND FLOW WAS RESTORED. THERE WAS NO REPORTED PATIENT INJURY AND THE PATIENT IS CURRENTLY REPORTED TO BE "OKAY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11024562 ↗InjuryOct 9, 2020Activation Failure; Activation Failure
IT WAS REPORTED THAT AFTER DEPLOYMENT OF THE FRED, THE MIDDLE OF THE STENT DID NOT COMPLETELY OPEN AROUND A CURVE IN THE VESSEL. THE STENT WAS RETRIEVED FROM THE PATIENT WITH A SNARE. THERE WAS NO REPORTED PATIENT INJURY. THE PATIENT IS CURRENTLY REPORTED TO BE "FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10659798 ↗InjuryApr 2, 2020Material Twisted/Bent; Activation Failure; Material Twisted/Bent; Activation Failure
IT WAS REPORTED THAT AFTER THE FRED HAD BEEN DEPLOYED IN THE SUPRACLINOID INTERNAL CAROTID ARTERY (ICA), PART OF THE STENT DID NOT APPEAR TO BE OPENED. A WIRE, MICROCATHETER, AND A GUIDE CATHETER WERE USED TO OPEN THE STENT. INTEGRILIN WAS ADMINISTERED TO THE PATIENT DURING THE PROCEDURE. UPON CLOSER INSPECTION POST-PROCEDURE, THE IMPLANTED STENT APPEARED TO BE TWISTED IN THE PROXIMAL SEGMENT. THERE WAS NO REPORTED PATIENT INJURY AND THE PATIENT IS REPORTED TO BE IN GOOD CONDITION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9918024 ↗
Other Intracranial aneurysm flow diverter devices
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page