FRED Flow Re-Direction Endoluminal Device, REF:… recallZ-2238-2023 · Microvention Inc. · initiated Jun 7, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Microvention Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Intracranial aneurysm flow diverter device may be manufactured with an incorrect inner stent length and may not perform as intended.
FDA root cause category: Process control
What the company is doing
On 6/7/23 recall notices were emailed to customers who were asked to stop use, quarantine, and return affected devices. Follow-up notices were mailed to customers on 6/16/23. Customers were asked to share the notice with all device users within the facility and network and share the notice with any organization where potentially affected devices have been transferred. In addition, customers were asked to complete and return the customer acknowledgement form. Inquiries can be directed to customer service at 800-990-8368.
Recall details
- Recall number
- Z-2238-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 7, 2023
- Posted by the FDA
- Jul 27, 2023
- Recalling firm
- Microvention Inc.
- Quantity
- 474
- Product code
- OUT
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, NY, FL, CA, PA, SC, TX, IL, MA, NJ, VA, GA, OH, IN, MI, KY, AL, MO, SD, LA, WA, HI, KS, MD, OK, DE, NC, MS and the countries of Taiwan, Chile, India, South Korea, Malaysia, Canada, France, Morocco, Belgium, Spain, Turkey, Poland, Russian Federation, Czech Republic, Slovenia, Germany, United Kingdom, El Salvador, United Arab Emirates, Bulgaria, Portugal, Brazil, Iran, Malta, Vietnam, South Africa, Jordan, Japan, Uruguay.
Product description
FRED Flow Re-Direction Endoluminal Device, REF: FRED3507-PMA, FRED3511, FRED3511-PMA, FRED3516-CA, FRED3516-PMA, FRED3524, FRED3524-CA, FRED3524-PMA, FRED3536-CA, FRED3536-PMA, FRED4007-PMA, FRED4012-CA, FRED4012-PMA, FRED4017-PMA, FRED4026, FRED4026-PMA, FRED4038-PMA, FRED4508, FRED4508-PMA, FRED4513-PMA, FRED4518-PMA, FRED4528, FRED4528-PMA, FRED4539, FRED4539-CA, FRED4539-PMA, FRED5009, FRED5009-PMA, FRED5014-PMA, FRED5019-PMA, FRED5029, FRED5029-CA, FRED5029-PMA, FRED5514-CA, FRED5514-PMA, FRED5526-CA, FRED5526-PMA, MV-F351127, MV-F352427, MV-F401227, MV-F401727, MV-F451327, MV-F500927, MV-F501427, MV-F501927, MV-F552627
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
Devices manufactured between June 1, 2020, and September 29, 2020. REF/UDI/Lot: FRED3507-PMA/00842429106655/200703556, 20070654P, 20072255L, 20072255N, 20073056A, 20073056B, 200807599, 20081453X, 20082656F, 20082656G, 20082852H, 20083152E; FRED3511/20073054C, 20091654Z, 200917511; FRED3511-PMA/00842429106662/20070754K, 200716562, 200721573, 20072252J, 20072255P, 20072255Q, 20080851B, 20082656H, 20082656J, 20090152R; FRED3516-CA/20081154U; FRED3516-PMA/00842429106679/20070654Q, 200716563, 20072158F, 200731555, 20080354B, 20081154W; FRED3524/00811425020982/200702553, 200706563, 20071557U, 200720563, 20081955R, 20091654T, 20091857N, 20092357X; FRED3524-CA/200924555; FRED3524-PMA/00842429106686/200716564, 200716565, 20072255U, 20072255V, 2007225AZ, 20073056C, 20073056Q, 20073056R, 20080459V, 20080558E, 20080853W, 2008115AC, 20081255W, 20081354Q, 200909564, 200909565, 200909664, 20091558Z, 20092156J, 20092166J; FRED3536-CA/20092859F, 20092955K; FRED3536-PMA/20070654R, 20071556Z, 200716566, 20072255W, 20072255X, 20080558G, 20081255X, 20090955W, 200915547, 200915591, 200916569, 200923539, 20092356W, 20092356X, 20092957Y; FRED4007-PMA/00842429106709/20072255Y, 20072255Z, 20072455H, 20073056Y, 20073056Z, 200730571, 20080558H, 20080558K, 20090255J, 20091555N; FRED4012-CA/20081255H; FRED4012-PMA/00842429106723/200703557, 200703575, 200730572, 200730573, 200806569, 20081957L, 20082656Q, 20092356Y, 20092356Z, 200929581; FRED4017-PMA/00842429106730/20070154P, 200702514, 200703558, 200730575, 200806589, 200812562, 20081354G, 200813554, 200901548, 200902548, 2009085AT, 2009095B2; FRED4026/20070754G, 20080556Z, 20081356V, 20081955T, 200928541M; FRED4026-PMA/00842429106747/20070452Q, 200730577, 20081453Y, 200923571, 200923572, 200927516, 200928532, 200928541, 20093055H, 20093065H; FRED4038-PMA/00842429106754/200722561, 200722562, 20082656R, 20083152F, 200831594, 200901595, 200902545, 200902546, 20090759V, 200909569, 20091656E, 20091659G, 2009175A1, 2009175A3, 20091953C, 20091953E, 20091953F, 20092156N, 20092156R, 2009225AX, 200923573, 200923574, 20092459J, 20092469J, 20092557J, 200926535, 20092956H, 20093056X, 200930586; FRED4508/00811425021057/200805571, 20080851A, 20081154R; FRED4508-PMA/00842429106761/200703559, 200722563, 200722564, 20073057C, 20073057E, 20082656V, 20082656W, 200831598, 2009015A1, 20091555Q; FRED4513-PMA/00842429106778/200702569, 20070256A, 200713549, 20073057F, 20073057G, 200807546, 200902562, 200923575, 200923576; FRED4518-PMA/00842429106785/200702597, 20070355A, 200721576, 20073057J, 20080851E, 200808535; FRED4528/20070653Z, 200716556, 20072258F; FRED4528-PMA/200902564, 200908524; FRED4539/200706541, 20071553E, 20092955L; FRED4539-CA/20092853Z, 20092957K; FRED4539-PMA/00842429106808/200722565, 200722566, 20080558Q, 20080558R, 200812568, 200812569, 20081957V, 20081957W, 20082656X, 20082656Y, 20082656Z, 20082954E, 20083152G, 20083154W, 200901594, 200902565, 200902566, 2009025AQ, 2009025CB, 200903594, 20090451X, 20090956C, 20091055L, 20091253U, 20091656F, 20091659F, 20091751N, 20091952B, 20091953A, 200922537, 200922584, 200922585, 200922684, 200923577, 200923578, 200923678, 20092553C, 20092553E, 200925571, 20092557H, 20092563C, 200926522, 20092652C, 20092858U, 20092868U, 200929583; FRED5009/20081254Z, 20081355F; FRED5009-PMA/00842429106815/20070256B, 200703574, 200722567, 20073057P, 20073057Q, 20073057R, 20080457X, 20080558T, 20080558U, 20081258J, 200902567; FRED5014-PMA/00842429106822/20070256E, 20070256F, 200706571, 20071553L, 20071651L, 200717545, 2007225B1, 200811596, 20082559G; FRED5019-PMA/00842429106839/20070754L, 200721577, 20073057T, 200826571, 20090152T, 200901593, 2009015A6, 2009025BZ; FRED5029/20073054E, 200812551; FRED5029-CA/20091655V, 20091665V, 20092156Q, 20092166Q; FRED5029-PMA/00842429106846/20070256J, 20070256K, 20070256L, 20070452X, 200713556, 2007135A3, 20081957Y, 200922539, 200923579, 20092357A, 200928542, 200929584; FRED5514-CA/20071655K, 2009085AU, 200922536; FRED5514-PMA/00842429106853/20070256N, 20070452Y, 20072258C, 20081957Z, 200819581, 20082655Z, 200826561, 200826661, 20082758U, 20090154H, 200902568, 200904552, 20090457F, 20090956G; FRED5526-CA/20092957L; FRED5526-PMA/00842429106860/20070254L, 20070256P, 20070256Q, 20070256R, 20071557R, 200819583, 200826573, 20092156L, 20092357B, 200929585; MV-F351127/20081358X, 20090954X, 20090956K; MV-F352427/200722M5U, 200730M6Q; MV-F401227/200730M72, 200806M69, 200819M7L, 200826M6Q, 200923M6Z; MV-F401727/200701M4P, 200901M48, 200909MB2; MV-F451327/200702M69, 200702M6A, 200713M49, 200730M7F, 200730M7G, 200807M46, 200923M75; MV-F500927/20090956N; MV-F501427/200702M6F, 200706M71, 200715M3L, 200716M1L, 200811M96; MV-F501927/200730M7T; MV-F552627/200702M4L, 200702M6Q
Devices named in this recall
20 devices in the FDA device registry whose identifier appears in the recall's code information.
- Stent FRED4007-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 death8 reports
- Stent FRED4012-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 death8 reports
- Stent FRED4017-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall8 reports
- Stent FRED3516-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall5 reports
- Stent FRED4508-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall5 reports
- Stent FRED5514-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall5 reports
- Stent FRED4518-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall4 reports
- Stent FRED3507-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 death3 reports
- Stent FRED5019-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall3 reports
- HydroCoil Helical 140205HES-VMicrovention, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION1 recall2 reports
- Stent FRED3511-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall2 reports
- Stent FRED4026-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 death2 reports
- Stent FRED4513-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall2 reports
- Stent FRED5014-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall2 reports
- Stent FRED3524-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 report
- Stent FRED5009-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 report
- Stent FRED5029-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 report
- Stent FRED5526-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 report
- Stent FRED4038-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall0 reports
- Stent FRED4539-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page