Sofia DA6115ST-GC-ODRecalls & reported problems
Microvention, Inc. · Sofia 6F-115cm STR · Class 2
All models of Sofia (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842429108444), received between Dec 14, 2023 and Jun 5, 2025 — classified as 2 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Intracranial Hemorrhage1
- Obstruction/Occlusion1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED IN A RESTORE CLINICAL STUDY, THE NEXT DAY FOLLOWING THE MECHANICAL THROMBECTOMY PROCEDURE THAT WAS PERFORMED USING WITH THE SOFIA ASPIRATION CATHETER, THE PATIENT EXPERIENCED SUBARCHNOID HEMORRHAGE, AND THIS EVENT WAS SUMMARIZED AS A LEFT POSTERIOR INSULA WEDGE-SHAPED HYPODENSITY, REPRESENTING ACUTE/SUBACUTE CVA. THERE WAS NO TREATMENT PROVIDED, AS THE PATIENT WAS ONLY OBSERVED. NIHSS SCORE WAS NOTED TO BE 3 AND MRS SCORE WAS AT 4. ALTHOUGH REQUESTED, THERE WAS NO OTHER PATIENT INFORMATION PROVIDED BY THE SITE AT THIS TIME. IT WAS NOTED THAT THE RELATIONSHIP TO SOFIA DEVICE IS CONSIDERED POSSIBLY, THE RELATIONSHIP TO MECHANICAL THROMBECTOMY PROCEDURE IS CONSIDERED DEFINITELY, THE RELATIONSHIP TO STUDY DISEASE IS CONSIDERED POSSIBLY, AND THE RELATIONSHIP TO CONCURRENT CONDITION/TREATMENT IS CONSIDERED POSSIBLY.
Patient problems: Intracranial Hemorrhage
MAUDE report 22155645 ↗InjuryDec 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE SOFAST STUDY THAT IMAGING TAKEN ONE DAY POST THROMBECTOMY REVEALED RE-OCCLUSION OF DISTAL LEFT M1 SEGMENT. FURTHER INTERVENTION VIA THROMBECTOMY WAS DECIDED TO BE FUTILE AND DECISION WAS MADE NOT TO RE-TREAT AND PLACED THE PATIENT ON MEDICATION ONLY. THE CLINICIAN NOTED THAT THE RE-OCCLUSION MAY HAVE OCCURRED DUE TO INTRACRANIAL ATHEROSCLEROSIS. THE EVENT WAS POSSIBLY RELATED TO THE STUDY DISEASE AND POSSIBLY TO CONCURRENT CONDITION. THE EVENT OUTCOME IS ONGOING.
Patient problems: Obstruction/Occlusion
MAUDE report 18332278 ↗
Other Catheter, intravascular, diagnostic devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page