Cervical Management Base Unit 5979-1Recalls & reported problems
Mizuho Orthopedic Systems, Inc. · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00842430106101), received between Jan 20, 2020 and Mar 31, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device Slipped” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Slipped1
- Loss of or Failure to Bond1
- Use of Device Problem1
- Use of Incorrect Control/Treatment Settings1
Patient problems
- Insufficient Information1
- Laceration(s)1
- No Known Impact Or Consequence To Patient1
- Paralysis1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMar 31, 2026Use of Device Problem
TRIOS SPINAL TABLE WITH CMBU (CERVICAL MANAGEMENT BASE UNIT) WAS UTILIZED FOR AN ANTERIOR-POSTERIOR FLIP PROCEDURE ON AN 80-YEAR-OLD PATIENT WEIGHING APPROXIMATELY 75-85 LBS. THE HOSPITAL PERSONNEL HAD NOT PERFORMED THIS TECHNIQUE IN APPROXIMATELY 10 YEARS. PRIOR TO THE PROCEDURE, THE INITIAL REPORTER RECOMMENDED RESCHEDULING TO ALLOW FOR IN-SERVICE TRAINING; THIS RECOMMENDATION WAS DECLINED BY THE ASSISTANT SURGEON. DURING SETUP, THE CMBU WAS ATTACHED WITH DIRECTION FROM THE INITIAL REPORTER. THE SURGEON TIGHTENED THE CONNECTION OF MAYFIELD SKULL CLAMP WITH CMBU DESPITE NOT BEING RECOMMENDED BY THE INITIAL REPORTER. THE INITIAL REPORTER INTERVENED A SECOND TIME TO CORRECT IMPROPER POSITIONING OF THE CHEST PLATE AND HIP/THIGH COMBINATION COMPONENTS BEFORE COMPRESSION WAS ATTEMPTED. DUE TO THE PATIENT'S SMALL FRAME, ADEQUATE COMPRESSION COULD NOT BE ACHIEVED. WHEN THE FLAT TOP WAS RAISED IN AN ATTEMPT TO COMPENSATE, THE MAYFIELD SKULL CLAMP AND CMBU WERE IN A LOCKED STATE, RESULTING IN DISPLACEMENT OF THE SKULL PINS POSTERIORLY AND CAUSING LACERATIONS TO THE PATIENT. THE PROCEDURE WAS HALTED. THE SPINAL TABLE WAS REMOVED, AND THE PATIENT WAS SAFELY TRANSFERRED TO A BED. THE TABLE TO
Patient problems: Laceration(s)
MAUDE report 24745395 ↗MalfunctionOct 11, 2022Use of Incorrect Control/Treatment Settings; Device Slipped; Use of Incorrect Control/Treatment Settings; Device Slipped
IT WAS REPORTED THAT THERE WAS A STARBURST SLIP ON A CERVICAL MANAGEMENT BASE UNIT DURING SURGERY.
Patient problems: Insufficient Information; Paralysis
MAUDE report 15584333 ↗MalfunctionJan 20, 2020Loss of or Failure to Bond; Loss of or Failure to Bond
CUSTOMER REPORTED SOME BROKEN PARTS WERE EPOXIED BACK TOGETHER, BUT BROKE AGAIN ALONG THE GLUE LINE. THERE WAS NO INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9606609 ↗
Other Table, Surgical With Orthopedic Accessories, Ac-Powered devices
- TS 7000 1604788Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall4 reports
- TS 7500 1717023Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall1 report
- TS 7500 1704695Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854085Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854087Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- ADVANCED CONTROL I-BASE, DC 5803Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered47 reports
- TRUMPF MEDICAL 1723633Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED43 reports
- TABLE, HANA, 110 V 6875Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered2 deaths36 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.