HANA LEG SPAR LIFT ASSIST, LEFT 6875-3900Recalls & reported problems
Mizuho Orthopedic Systems, Inc. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842430109010), received between Apr 23, 2020 and Jun 9, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Positioning Problem1
- Unintended Movement1
Patient problems
- Injury1
- No Code Available1
- Perforation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2020Device Dislodged or Dislocated; Unintended Movement; Device Dislodged or Dislocated; Unintended Movement
IT WAS REPORTED TABLE WOULD NOT HOLD ITS POSITION, THEREFORE THE PATIENT WAS NOT POSITIONED CORRECTLY. THIS CAUSE REAMING THOUGH THE SIDE WALL OF THE FEMUR.
Patient problems: Injury; Perforation
MAUDE report 10134561 ↗MalfunctionApr 23, 2020Positioning Problem
THE HANA TABLE WOULD NOT HOLD ITS POSITION, THEREFORE THE PATIENT WAS NOT POSITIONED CORRECTLY. THIS CAUSED REAMING THROUGH THE SIDE WALL OF THE FEMUR. THE GENTLEMAN WHO WAS CALLED TO SERVICE THE BED SEEMED TO THINK THAT THE CAUSTIC NATURE OF THE CLEANER MAY HAVE CONTRIBUTED TO THE BREAKDOWN OF THE INNER MECHANISMS USED TO LOCK THE ARTICULATING PIECES IN PLACE. MANUFACTURER RESPONSE FOR TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED, HANA LEG SPAR LIFT ASSIST, LEFT (PER SITE REPORTER). SEE ORIGINAL DESCRIPTION.
Patient problems: No Code Available
MAUDE report 9990118 ↗
Other Table, Surgical With Orthopedic Accessories, Ac-Powered devices
- TS 7000 1604788Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall4 reports
- TS 7500 1717023Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall1 report
- TS 7500 1704695Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854085Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854087Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- ADVANCED CONTROL I-BASE, DC 5803Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered47 reports
- TRUMPF MEDICAL 1723633Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED43 reports
- TABLE, HANA, 110 V 6875Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered2 deaths36 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page