CERVICAL TRACTION VECTOR ADJUSTER 7803-790Recalls & reported problems
Mizuho Orthopedic Systems, Inc. · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00842430111679), received between Apr 1, 2021 and Apr 4, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Device Slipped1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 4, 2023Device Slipped; Device Slipped
PATIENT WAS HAVING A POSTERIOR CERVICAL FUSION USING GARDNER WELLS TONGS AND TRIOS VECTOR TRACTION. TRIOS VECTOR TRACTION SLIPPED MID CASE AND PATIENTS HEAD FELL AND WAS CAUGHT BY THE HORSESHOE HEADREST. PATIENT WAS NOT INJURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16677760 ↗MalfunctionApr 27, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED A BRAND-NEW TRACTION VECTOR CONNECTED TO 15 LBS OF WEIGHT AND GW TONGS ATTACHED TO PATIENT'S HEAD POPPED OFF UPPER TOWER OF TRIOS TABLE DURING A POST CERVICAL PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11733719 ↗MalfunctionApr 1, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
CUSTOMER REPORTED THAT TRACTION VECTOR POPPED OFF TOWER IN THE MIDDLE OF A CERVICAL CASE. THE SURGEON APPLIED THE VECTOR HIMSELF AND USES THIS ATTACHMENT FREQUENTLY. THE PATIENT WAS NOT INJURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11602149 ↗
Other Table, Surgical With Orthopedic Accessories, Ac-Powered devices
- TS 7000 1604788Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall4 reports
- TS 7500 1717023Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall1 report
- TS 7500 1704695Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854085Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854087Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- ADVANCED CONTROL I-BASE, DC 5803Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered47 reports
- TRUMPF MEDICAL 1723633Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED43 reports
- TABLE, HANA, 110 V 6875Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered2 deaths36 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗