Skip to content

Syva® Emit® Disopyramide Assay 10445370Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Syva® Emit® Disopyramide Assay · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 26, 2026Class IIIOpen, Classified

    Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.

    Z-2880-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other ENZYME IMMUNOASSAY, LIDOCAINE devices

All ENZYME IMMUNOASSAY, LIDOCAINE devices →