Skip to content

Syva Emit Disopyramide Assay recallZ-2880-2026 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 26, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.

FDA root cause category: Under Investigation by firm

What the company is doing

On 6/26/2026, Siemens issued Urgent Medical Device Correction letters to all affected customers informing them of the issue and providing instructions. Customers are instructed to refer to the SDS which were created and provided with the Urgent Medical Device Corrections and the Urgent Field Safety Notices specifically for Syva Emit Disopyramide Assay Lot U1 for information regarding handling and disposal.

Recall details

Recall number
Z-2880-2026
Status
Open, Classified
Initiated by the company
Jun 26, 2026
Posted by the FDA
Aug 4, 2026
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
8 US; 44 OUS
Product code
KLR
FDA submissions
K801204
Distribution
Worldwide - US Nationwide distribution in the states of CA and VA. The country of Japan.

Product description

Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL

Code information (models, lots, serial numbers)

UDI-DI 00842768001406 Lot U1

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates