Dade® Thrombin Reagent 10445721Recalls & reported problems

Siemens Healthcare Diagnostics Products GmbH · Determination of fibrinogen in plasma and to accelerate coagulation of anticoagulated samples for immunohematology studies · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842768003950), received between Jul 25, 2024 and Jul 25, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Low Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2024Low Test Results; Low Test Results

    THE CUSTOMER REPORTED THE OBSERVATION OF FALSELY DEPRESSED FIBRINOGEN RESULTS ON ONE PATIENT MEASURED WITH THE SIEMENS DADE THROMBIN REAGENT ON A SYSMEX CS-5100 SYSTEM (LEGAL MANUFACTURER: SYSMEX) COMPARED TO THE HISTORICAL FIBRINOGEN RESULTS OBTAINED ON AN ALTERNATE METHOD. THE PATIENT SAMPLE WAS PROCESSED IN 1:1 AND 4:1 DILUTIONS. FOR BOTH DILUTION RESULTS, THE INSTRUMENT GENERATED A FLAG AND DID NOT DISPLAY NUMERICAL RESULTS. HOWEVER, THE USER IGNORED OR MISINTERPRETED THE MESSAGE OR FLAG AND REPORTED THE RESULTS (AS <0.3 G/L) TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULTS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED FIBRINOGEN RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19826720 ↗

Other SYSTEM, FIBRINOGEN DETERMINATION devices