Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842768003950), received between Jul 25, 2024 and Jul 25, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED THE OBSERVATION OF FALSELY DEPRESSED FIBRINOGEN RESULTS ON ONE PATIENT MEASURED WITH THE SIEMENS DADE THROMBIN REAGENT ON A SYSMEX CS-5100 SYSTEM (LEGAL MANUFACTURER: SYSMEX) COMPARED TO THE HISTORICAL FIBRINOGEN RESULTS OBTAINED ON AN ALTERNATE METHOD. THE PATIENT SAMPLE WAS PROCESSED IN 1:1 AND 4:1 DILUTIONS. FOR BOTH DILUTION RESULTS, THE INSTRUMENT GENERATED A FLAG AND DID NOT DISPLAY NUMERICAL RESULTS. HOWEVER, THE USER IGNORED OR MISINTERPRETED THE MESSAGE OR FLAG AND REPORTED THE RESULTS (AS <0.3 G/L) TO THE PHYSICIAN(S), WHO QUESTIONED THE RESULTS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED FIBRINOGEN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19826720 ↗
Other SYSTEM, FIBRINOGEN DETERMINATION devices
- STA® - Fibrinogen 5 00674Diagnostica Stago · SYSTEM, FIBRINOGEN DETERMINATION1 recall0 reports
- Dade® Thrombin Reagent 10445720Siemens Healthcare Diagnostics Products GmbH · SYSTEM, FIBRINOGEN DETERMINATION1 report
- Dade® Owren's Veronal Buffer 10445724Siemens Healthcare Diagnostics Products GmbH · SYSTEM, FIBRINOGEN DETERMINATION1 report
- Multifibren® U 10446691Siemens Healthcare Diagnostics Products GmbH · SYSTEM, FIBRINOGEN DETERMINATION1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page