Skip to content

N Latex CDT Kit 10445997Recalls & reported problems

Siemens Healthcare Diagnostics Products GmbH · Quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum on BN II and BN ProSpec® Systems · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 2, 2020Class IITerminated

    Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring

    Z-1790-2020 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. May 31, 2018Class IITerminated

    Siemens Healthcare Diagnostics has observed a negative bias for the carbohydrate-deficient transferrin (CDT) measurement when using the affected N Latex CDT lots in comparison to the HPLC method. CDT results in absolute concentrations may be influenced by patients transferrin levels and, therefore, results are reported as ratio of CDT to total transferrin, called %CDT. As the transferrin determination with N Antiserum to Human Transferrin is not affected, the observed effect leads to calculated %CDT values that show a negative bias of approximately 15% compared to the HPLC method. This could result in a shift of weak positive patient samples into the reference range of 1.19 2.47 %CDT which was derived from a study population of healthy adults. There are no control or calibrator materials that would catch this bias.

    Z-2931-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page