Siemens N Latex CDT Kit- IVD quantitative determination of… recallZ-1790-2020 · CSL Behring GmbH · initiated Mar 2, 2020
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CSL Behring GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter PP-20-002.A.US to the 5 affected customers in the U.S. on March 5, 2020 via FedEx overnight delivery. ( Urgent Field Corrective Action (UFCA) PP20-002.A.OUS was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities). Letter states reason for recall, health risk, and action to take by the Customer Please review this letter with your Medical Director. Discontinue use of and discard the kit lots listed in Table 1. Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-1790-2020
- Status
- Terminated
- Initiated by the company
- Mar 2, 2020
- Recalling firm
- CSL Behring GmbH
- Quantity
- 297 PCK US; 1247 PCK OUS
- Product code
- NAO
- FDA submissions
- K060677
- Distribution
- IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan
Product description
Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997
Code information (models, lots, serial numbers)
US Lot number: 49822 UDI: 842768018534 OUS: Lot numbers: 49729, 49849, 49985, 50084 Exp. Date: 2020-04-11
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- N Latex CDT Kit 10445997Siemens Healthcare Diagnostics Products GmbH · Test,carbohydrate deficient transferrin2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page