INNOVANCE® D-Dimer 10445981Recalls & reported problems
Siemens Healthcare Diagnostics Products GmbH · For the quantitative determination of cross-linked fibrin degradation products (D-dimers) in human plasma on Siemens Sysmex® Coagulation Systems · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00842768022814), received between Nov 5, 2018 and Jun 30, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings3
- Low Test Results3
- Off-Label Use2
- False Negative Result1
- Use of Device Problem1
Patient problems
- No Known Impact Or Consequence To Patient7
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 30, 2025False Negative Result; False Negative Result
THE CUSTOMER REPORTED THE OBSERVATION OF A DISCORDANT NEGATIVE D-DIMER PATIENT SAMPLE RESULT OBTAINED ON PATIENT SAMPLE MEASURED WITH INNOVANCE D-DIMER ON CA-660 INSTRUMENT COMPARED TO REPEAT TESTING. THE DISCORDANT NEGATIVE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22355954 ↗MalfunctionMar 1, 2023Use of Device Problem; Use of Device Problem
A CUSTOMER REPORTED THAT A PHYSICIAN MISINTERPRETED AN INNOVANCE D-DIMER RESULT OF 0.93 MCG/ML WITH A REFERENCE RANGE OF 0.19 TO 1.12 MCG/ML FROM THE SYSMEX CS-2500 SYSTEM AS A RESULT WITHIN THE NORMAL RANGE, INSTEAD OF ABOVE THE CUT-OFF FOR THE EXCLUSION OF VENOUS THROMBOEMBOLISM (VTE). AS A RESULT, THE PHYSICIAN(S) SENT THE PATIENT HOME WITHOUT FURTHER TESTING. AFTER REALIZING THE MISINTERPRETATION, THE PHYSICIAN(S) RE-EVALUATED THE CASE AND DETERMINED THAT THE PATIENT WAS NOT AT RISK OF VTE. THIS REPORT IS BEING FILED IN AN ABUNDANCE OF CAUTION. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16460696 ↗MalfunctionAug 13, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A D-DIMER RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A SYSMEX CA-660 SYSTEM USING INNOVANCE D-DIMER REAGENT. THE CUSTOMER RERAN THE SAMPLE FOR D-DIMER USING THE D-DIMER+ (DDI+) APPLICATION WITH THE SAME SYSTEM AND REAGENT, RECOVERING LOWER. A SECOND SAMPLE FROM THE SAME PATIENT WAS RUN FOR D-DIMER USING THE SAME SYSTEM AND REAGENT, RECOVERING HIGHER. THE REDRAWN SAMPLE WAS REPEATED AGAIN FOR D-DIMER USING THE DDI+ APPLICATION WITH THE SAME SYSTEM AND REAGENT, RECOVERING LOWER. THE HIGHER D-DIMER RESULT WAS REPORTED TO THE PHYSICIAN(S). IT IS UNKNOWN WHICH D-DIMER RESULT WAS CONSIDERED CORRECT. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT D-DIMER RESULT(S).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10404576 ↗MalfunctionJul 20, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A DISCORDANT D-DIMER RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A SYSMEX CA-660 SYSTEM USING INNOVANCE D-DIMER REAGENT. THE DISCORDANT RESULT WAS REPORTED TO THE PHYSICIAN(S). PRIOR TO OBTAINING THE DISCORDANT RESULT, THE SAMPLE WAS RUN FOR D-DIMER TWICE ON THE SYSMEX CA-660 SYSTEM USING INNOVANCE D-DIMER REAGENT, BUT THE SYSTEM DID NOT YIELD NUMERIC RESULTS AND GENERATED MEASUREMENT ERROR FLAGS. THE FOLLOWING DAY, THE SAME SAMPLE WAS REPEATED FOR D-DIMER ON THE SAME SYSMEX CA-660 SYSTEM USING INNOVANCE D-DIMER REAGENT, RESULTING LOWER. THIS RESULT WAS ALSO CONSIDERED DISCORDANT AND WAS NOT REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT D-DIMER RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10295865 ↗MalfunctionJul 20, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A DISCORDANT D-DIMER RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A SYSMEX CA-660 SYSTEM USING INNOVANCE D-DIMER REAGENT. THE DISCORDANT RESULT WAS REPORTED TO THE PHYSICIAN(S). PRIOR TO OBTAINING THE DISCORDANT RESULT, THE SAMPLE WAS RUN FOR D-DIMER TWICE ON THE SYSMEX CA-660 SYSTEM USING INNOVANCE D-DIMER REAGENT, BUT THE SYSTEM DID NOT YIELD NUMERIC RESULTS AND GENERATED MEASUREMENT ERROR FLAGS. THE FOLLOWING DAY, THE SAME SAMPLE WAS REPEATED FOR D-DIMER ON THE SAME SYSMEX CA-660 SYSTEM USING INNOVANCE D-DIMER REAGENT, RESULTING LOWER. THIS RESULT WAS ALSO CONSIDERED DISCORDANT AND WAS NOT REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT D-DIMER RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10295862 ↗MalfunctionMay 6, 2020Low Test Results; Low Test Results
DISCORDANT, FALSELY LOW D-DIMER RESULTS WERE OBTAINED ON A PATIENT SAMPLE ON TWO DIFFERENT SYSMEX CA-1500 SYSTEMS USING INNOVANCE D-DIMER REAGENT. WHEN THE PATIENT SAMPLE WAS INITIALLY RUN FOR D-DIMER ON A SYSMEX CA-1500 SYSTEM USING INNOVANCE D-DIMER REAGENT, THE SYSTEM GENERATED A "RANGE OVER" ERROR CODE AND DID NOT GIVE A RESULT. THE SAMPLE WAS REPEATED FOR D-DIMER ON THE SAME SYSTEM WITH THE SAME REAGENT, RECOVERING ABOVE THE LAB'S D-DIMER CUTOFF. THIS RESULT WAS CONSIDERED CORRECT AND WAS REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS THEN REPEATED FOR D-DIMER ON THE SAME SYSTEM WITH THE SAME REAGENT, RECOVERING BELOW THE CUTOFF. THIS RESULT WAS CONSIDERED DISCORDANT AND WAS NOT REPORTED TO THE PHYSICIAN(S). THE FOLLOWING DAY, THE SAME SAMPLE WAS REPEATED FOR D-DIMER ON AN ALTERNATE SYSMEX CA-1500 SYSTEM USING D-DIMER REAGENT, WHICH ALSO RECOVERED BELOW THE CUTOFF. THIS RESULT WAS ALSO CONSIDERED DISCORDANT AND WAS NOT REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY LOW D-DIMER RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10028727 ↗MalfunctionMay 6, 2020Low Test Results; Low Test Results
DISCORDANT, FALSELY LOW D-DIMER RESULTS WERE OBTAINED ON A PATIENT SAMPLE ON TWO DIFFERENT SYSMEX CA-1500 SYSTEMS USING INNOVANCE D-DIMER REAGENT. WHEN THE PATIENT SAMPLE WAS INITIALLY RUN FOR D-DIMER ON A SYSMEX CA-1500 SYSTEM USING INNOVANCE D-DIMER REAGENT, THE SYSTEM GENERATED A "RANGE OVER" ERROR CODE AND DID NOT GIVE A RESULT. THE SAMPLE WAS REPEATED FOR D-DIMER ON THE SAME SYSTEM WITH THE SAME REAGENT, RECOVERING ABOVE THE LAB'S D-DIMER CUTOFF. THIS RESULT WAS CONSIDERED CORRECT AND WAS REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS THEN REPEATED FOR D-DIMER ON THE SAME SYSTEM WITH THE SAME REAGENT, RECOVERING BELOW THE CUTOFF. THIS RESULT WAS CONSIDERED DISCORDANT AND WAS NOT REPORTED TO THE PHYSICIAN(S). THE FOLLOWING DAY, THE SAME SAMPLE WAS REPEATED FOR D-DIMER ON AN ALTERNATE SYSMEX CA-1500 SYSTEM USING D-DIMER REAGENT, WHICH ALSO RECOVERED BELOW THE CUTOFF. THIS RESULT WAS ALSO CONSIDERED DISCORDANT AND WAS NOT REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY LOW D-DIMER RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10028723 ↗MalfunctionMay 29, 2019Off-Label Use; Off-Label Use
A DISCORDANT, FALSELY LOW D-DIMER RESULT WAS OBTAINED ON A (B)(6) FEMALE PATIENT USING INNOVANCE D-DIMER KIT LOT 48736 ON A SYSMEX CA-660 (SERIAL NUMBER: (B)(4)) SYSTEM. THE PATIENT UNDERWENT A SURGERY ABOUT A WEEK PRIOR TO THE DATE OF THE OBTAINED DISCORDANT RESULT AND THE REASON IS UNKNOWN. THE DISCORDANT RESULT WAS OBTAINED USING A BECTON-DICKINSON (BD) 1.8 ML COAGULATION TUBE; THE SAMPLE WAS REPEATED TWICE WITH EACH REPEAT BEING FLAGGED BY THE INSTRUMENT AND NON-NUMERICAL VALUE OBTAINED. THE INITIAL, NON-FLAGGED, RESULT WAS REPORTED TO THE PHYSICIAN(S) AND IT WAS QUESTIONED. THE PATIENT WAS REDRAWN IN A BD 2.7 ML COAGULATION TUBE FOR ADDITIONAL REPEAT TESTING AND THE RESULTS OBTAINED WERE HIGHER, WHICH MATCHED THE PATIENT CLINICAL HISTORY. A CORRECTED REPORTED WAS ISSUED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY LOW D-DIMER RESULT OBTAINED ON THE SYSMEX CA-660 SYSTEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8651273 ↗MalfunctionNov 5, 2018Off-Label Use; Low Test Results; Off-Label Use; Low Test Results
A DISCORDANT, FALSELY LOW D-DIMER RESULT WAS OBTAINED ON A PATIENT SAMPLE WHEN THE SAMPLE WAS DILUTED AND REPEATED ON A SYSMEX CA-1500 SYSTEM, USING THE INNOVANCE D-DIMER REAGENT. WHEN THE SAMPLE WAS INITIALLY RUN, A HIGHER D-DIMER RESULT WAS OBTAINED ON THE PATIENT SAMPLE. THE DISCORDANT RESULT WAS REPORTED TO THE PHYSICIAN(S) AND RECALLED BEFORE THE PHYSICIAN(S) SAW THE RESULT. THE SAMPLE WAS REPEATED MULTIPLE TIMES ON THE SAME SYSTEM UNDER DIFFERENT SAMPLE IDS, RESULTING HIGHER. THE REPEAT RESULT OF 4.09 WAS REPORTED, AS THE CORRECT RESULT, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY LOW D-DIMER RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8039059 ↗
Other FIBRINOGEN AND FIBRIN SPLIT PRODUCTS, ANTIGEN, ANTISERUM, CONTROL devices
- STA® - Liatest® D-Di 00515Diagnostica Stago · FIBRINOGEN AND FIBRIN SPLIT PRODUCTS, ANTIGEN, ANTISERUM, CONTROL2 recalls0 reports
- VIDAS® D-Dimer Exclusion II ™ 30455-01Biomerieux SA · Fibrinogen And Fibrin Split Products, Antigen, Antiserum, Control1 recall2 reports
- INNOVANCE® D-Dimer 10445982Siemens Healthcare Diagnostics Products GmbH · FIBRINOGEN AND FIBRIN SPLIT PRODUCTS, ANTIGEN, ANTISERUM, CONTROL8 reports
- Alere Triage® 98100Alere San Diego, Inc. · Fibrinogen And Fibrin Split Products, Antigen, Antiserum, Control4 reports
- INNOVANCE® D-Dimer Controls 10446006Siemens Healthcare Diagnostics Products GmbH · FIBRINOGEN AND FIBRIN SPLIT PRODUCTS, ANTIGEN, ANTISERUM, CONTROL2 reports
- Triage 98100Quidel Cardiovascular Inc. · FIBRINOGEN AND FIBRIN SPLIT PRODUCTS, ANTIGEN, ANTISERUM, CONTROL1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page