Skip to content

INNOVANCE® D-Dimer Controls 10446006Recalls & reported problems

Siemens Healthcare Diagnostics Products GmbH · INNOVANCE® D-Dimer Control 1 and Control 2 are assayed, normal and pathological level, intralaboratory quality controls for assessment of precision and analytical bias in the quantitative determination of D-dimer on Siemens and Sysmex® Systems · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842768023668), received between Oct 6, 2017 and Oct 6, 2017 — classified as 2 malfunctions.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 6, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CUSTOMER REPORTED HIGH, OUT OF RANGE RECOVERY OF D-DIMER CONTROL LEVEL 2 ON THE SYSMEX CA-1500 SYSTEM AFTER CHANGING QUALITY CONTROL (QC) LOTS ON (B)(6) 2017. D-DIMER CONTROL QC LEVEL 1 WAS ALWAYS WITHIN THE TABLE OF ASSIGNED VALUES (TAV) RANGE. A PATIENT SAMPLE WAS RUN AND THE RESULT WAS REPORTED TO THE PHYSICIAN WHILE THE LEVEL 2 QC WAS OUT OF RANGE. THE PHYSICIAN DID NOT QUESTION THE RESULT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE POTENTIALLY DISCORDANT D-DIMER RESULT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6922082 ↗
  • MalfunctionOct 6, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A CUSTOMER REPORTED HIGH, OUT OF RANGE RECOVERY OF D-DIMER CONTROL LEVEL 2 ON THE SYSMEX CA-1500 SYSTEM AFTER CHANGING QUALITY CONTROL (QC) LOTS ON (B)(6) 2017. D-DIMER CONTROL LEVEL 1 WAS ALWAYS WITHIN THE TABLE OF ASSIGNED VALUES (TAV) RANGE. PATIENT SAMPLES WERE RUN AND RESULTS WERE REPORTED TO THE PHYSICIAN(S) WHILE THE LEVEL 2 QC WAS OUT OF RANGE. THE PHYSICIAN(S) DID NOT QUESTION THE RESULTS. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE POTENTIALLY DISCORDANT D-DIMER RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6922077 ↗

Other FIBRINOGEN AND FIBRIN SPLIT PRODUCTS, ANTIGEN, ANTISERUM, CONTROL devices

All FIBRINOGEN AND FIBRIN SPLIT PRODUCTS, ANTIGEN, ANTISERUM, CONTROL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page