Skip to content

Dimension Vista® Flex® reagent cartridge PROG 10461743Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · LOCI Progesterone Flex® reagent cartridge, 320 Tests · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842768026478), received between Feb 18, 2021 and Feb 18, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2021High Test Results; High Test Results

    A DISCORDANT, FALSELY ELEVATED PROGESTERONE (PROG) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION VISTA 500 SYSTEM. THE RESULT WAS REPORTED AND QUESTIONED BY THE PHYSICIAN. THE SAME SAMPLE WAS REPROCESSED ON MULTIPLE LATER DATES AND SIMILAR RECOVERIES WERE OBTAINED. THE SAME SAMPLE WAS PROCESSED BY TWO ALTERNATE METHODOLOGIES AND LOWER, BELOW ASSAY RANGE RESULTS WERE OBTAINED. THE LOWER RESULTS WERE CONSIDERED CORRECT BY CUSTOMER, CONSISTENT WITH PATIENT HISTORY. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCREPANT PROGESTERONE RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11345928 ↗

Other RADIOIMMUNOASSAY, PROGESTERONE devices

All RADIOIMMUNOASSAY, PROGESTERONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page