Masimo Corporation · Radius-7 Wearable Pulse CO-Oximeter · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Masimo Radius-7 is a wearable pulse CO-oximeter that monitors oxygen and related blood levels. The 23 reports naming it mostly describe the device sensing intermittently, losing its connection, freezing displays, or shutting down unexpectedly; most reported no harm to the patient, though one report involved a death and no injuries were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00843997009249), received between Nov 21, 2017 and Jan 19, 2024 — classified as 1 death, 22 malfunctions.

The most frequently reported specific device problem is “Sensing Intermittently” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Sensing Intermittently7
  • Loose or Intermittent Connection5
  • Display or Visual Feedback Problem3
  • Inappropriate or Unexpected Reset2
  • Incorrect Measurement2
  • Unexpected Shutdown2
  • Disconnection1
  • No Audible Alarm1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • No Consequences Or Impact To Patient8
  • Death1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RADIUS-7 SHUTS DOWN. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18547471 ↗
  • MalfunctionJul 21, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE DEVICE SHUTS OFF AFTER 10 MINUTES. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15080272 ↗
  • MalfunctionApr 19, 2022Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE CUSTOMER REPORTED THE MONITOR DISPLAY WAS FROZEN. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14149412 ↗
  • MalfunctionApr 19, 2022Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE CUSTOMER REPORTED THE MONITOR DISPLAY WAS FROZEN. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14149383 ↗
  • MalfunctionApr 19, 2022Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE CUSTOMER REPORTED THE MONITOR DISPLAY WAS FROZEN. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14149344 ↗
  • MalfunctionJan 4, 2022Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THE MODULE WORKS INTERMITTENTLY. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13155652 ↗
  • MalfunctionDec 14, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    THE CUSTOMER REPORTED INTERMITTENT CONNECTION. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13006418 ↗
  • MalfunctionDec 14, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    THE CUSTOMER REPORTED INTERMITTENT CONNECTION. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13006379 ↗
  • MalfunctionDec 8, 2021Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THE DEVICE IS INTERMITTENTLY DISCONNECTING. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12965030 ↗
  • MalfunctionJun 24, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    THE CUSTOMER REPORTED THE RADIUS-7 INTERMITTENTLY DISCONNECTS FROM THE MODULE. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12063152 ↗
  • MalfunctionJun 10, 2021Disconnection; Disconnection

    THE CUSTOMER REPORTED THE DEVICE GLITCHES AND PERIODICALLY DISCONNECTS. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11975359 ↗
  • MalfunctionApr 9, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    THE CUSTOMER REPORTED THE DEVICE LOSES CONNECTION WITH BOARD. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11641171 ↗
  • MalfunctionApr 1, 2021Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    THE CUSTOMER REPORTED THAT THE DEVICE SHUTS DOWN INTERMITTENTLY. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11603076 ↗
  • MalfunctionApr 1, 2021Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    THE CUSTOMER REPORTED THAT THE DEVICE SHUTS DOWN INTERMITTENTLY. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11603068 ↗
  • MalfunctionMar 4, 2021Incorrect Measurement; Incorrect Measurement

    THE CUSTOMER REPORTED INACCURATE READINGS. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11414223 ↗
  • MalfunctionJan 8, 2021Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THE DEVICE WILL NOT HOLD A SIGNAL. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11140548 ↗
  • MalfunctionJan 8, 2021Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THE DEVICE WILL NOT HOLD A SIGNAL. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11140530 ↗
  • MalfunctionMay 11, 2020Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THAT THE DEVICE WAS INTERMITTENTLY DISCONNECTS SPO2. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10043683 ↗
  • MalfunctionOct 7, 2019Loose or Intermittent Connection; Loose or Intermittent Connection

    THE CUSTOMER REPORTED DROPPED CONNECTION WHEN SLIGHTLY MOVED. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9162339 ↗
  • MalfunctionMay 22, 2019Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED INTERMITTENT OPERATION OF THE RADIUS-7 MODULE WHEN THE SENSOR IS PLUGGED IN. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8634668 ↗
  • MalfunctionApr 23, 2018Unable to Obtain Readings; Sensing Intermittently; Unable to Obtain Readings; Sensing Intermittently

    THE CUSTOMER REPORTED THEY STARTED THE TEST AND IN A FEW SECONDS THE DEVICE STOPPED GIVING A READING, THE PULSE CURVE DRASTICALLY DROPPED, AS DID THE SPO2 CURVE AND IP. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7449930 ↗
  • DeathJan 19, 2018No Audible Alarm; No Audible Alarm

    IT WAS REPORTED: "NURSING STAFF AND A PHYSICIAN WERE PRESENT AT THE PATIENT BEDSIDE WHEN THE PATIENT WAS FOUND WITHOUT A PULSE. THE TELEMETRY TECH FROM PCU CALLED TO SAY THE PATIENT WAS IN ASYSTOLE AT THAT TIME. A RAPID RESPONSE WAS ALREADY UNDERWAY. STAFF STATED THE MONITOR DID NOT ALARM IN THE ROOM NOR WAS THE STAFF PAGER INITIATED." PER THE CUSTOMER, THE PATIENT WAS TRANSFERRED TO THE ICU AS THE SAFETYNET SYSTEM AND BEDSIDE MONITORS FAILED TO NOTIFIY CLINICIANS REGARDING THE PATIENT CONDITION. THE PATIENT PASSED AWAY IN THE ICU.

    Patient problems: Death

    MAUDE report 7208569 ↗
  • MalfunctionNov 21, 2017Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED: ¿SPO2 WAS OFF WHEN COMPARED TO A RAD 5V BY AROUND 10-15%. RADIUS 7 DISCONNECT AS WELL" NO CONSEQUENCES OR IMPACT TO PATIENT WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7050882 ↗