RD SET™ MP-12 RD SET to IVM/MMS 4083Recalls & reported problems
Masimo Corporation · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00843997009782), received between Nov 3, 2016 and Dec 2, 2021 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Device Stops Intermittently” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Stops Intermittently3
- Incorrect Measurement3
- Sensing Intermittently3
- Device Operates Differently Than Expected1
- High Readings1
- Inappropriate or Unexpected Reset1
- Sparking1
Patient problems
- No Consequences Or Impact To Patient7
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2021Sensing Intermittently; Sensing Intermittently
THE CUSTOMER REPORTED THE SENSOR WAS INTERMITTENT. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12919703 ↗MalfunctionAug 9, 2021High Readings; High Readings
THE CUSTOMER REPORTED HEART RATE READINGS WERE HIGH. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12298070 ↗MalfunctionAug 20, 2019Incorrect Measurement; Incorrect Measurement
THE CUSTOMER REPORTED WHEN THE CABLE IS CONNECTED AND BEING USED THE SPO2 READINGS WERE STAYING IN THE 80S EVEN WHEN THE SENSOR WAS MOVED TO DIFFERENT SITES THE READINGS REMAINED IN THE 80S. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8908943 ↗MalfunctionOct 12, 2018Sparking; Sparking
THE CUSTOMER REPORTED THAT WHEN A SENSOR WAS PLUGGED INTO THE CABLE, THERE WAS A SPARK, AND THE CABLE HAS BURN MARKS ON IT. NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7962341 ↗MalfunctionFeb 16, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
THE CUSTOMER REPORTED: "CABLE SPLITTING AT CASTING OF SENSOR CONNECTION END; THEREFORE CAUSING INTERMITTENT READINGS, DROPPING OUT SIGNALS, AND ISSUES OVERALL WITH MONITORING EFFECTIVELY AND ACCURATELY." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7279229 ↗MalfunctionFeb 16, 2018Sensing Intermittently; Sensing Intermittently
THE CUSTOMER REPORTED: "RD PHILIPS CABLE PART NUMBER 4083 SPLITTING AT CASING OF SENSOR CONNECTION END; THEREFORE CAUSING INTERMITTENT READINGS, DROPPING OUT SIGNALS, AND CAUSING ISSUES OVERALL WITH MONITORING EFFECTIVELY AND ACCURATELY." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7279226 ↗MalfunctionFeb 16, 2018Sensing Intermittently; Sensing Intermittently
THE CUSTOMER REPORTED: "RD PHILIPS CABLE PART NUMBER 4083 SPLITTING AT CASING OF SENSOR CONNECTION END; THEREFORE CAUSING INTERMITTENT READINGS, DROPPING OUT SIGNALS, AND CAUSING ISSUES OVERALL WITH MONITORING EFFECTIVELY AND ACCURATELY." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7279225 ↗MalfunctionSep 22, 2017Device Stops Intermittently; Device Stops Intermittently
THE CUSTOMER REPORTED CABLE WORKS INTERMITTENTLY. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6888960 ↗MalfunctionSep 22, 2017Device Stops Intermittently; Device Stops Intermittently
THE CUSTOMER REPORTED CABLE WORKS INTERMITTENTLY. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6888955 ↗MalfunctionJun 30, 2017Device Stops Intermittently; Device Stops Intermittently
THE CUSTOMER REPORTED CABLE WORKS INTERMITTENTLY. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6683059 ↗MalfunctionMar 24, 2017Incorrect Measurement; Incorrect Measurement
THE CUSTOMER REPORTED THE VALUE OF SP2 WAS LOW (45-90%). THE EVENT WAS NOT RESOLVED WITH THE REPLACEMENT. NO PATIENT IMPACT OR CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6433992 ↗MalfunctionDec 2, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT LOOSE CONNECTION AT MONITOR PLUG. IF YOU MOVE THE CABLE AT THE ROUND PLUG, THERE IS SPORADICALLY NO MORE CURVE. PER THE CUSTOMER, THERE WAS A MEASUREMENT IN THE BEGINNING, BUT FURTHER ON THE ARTIFACTS INCREASED, THE MEASUREMENT DISAPPEARED. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6141955 ↗MalfunctionNov 3, 2016Incorrect Measurement; Incorrect Measurement
THE CUSTOMER REPORTED SEEING AN SPO2 OF 39% ON THE PATIENT WHICH DID NOT MATCH THE CLINICAL PICTURE WHEN USING A PHILIPS MP 50 WITH FAST VIA A MMS MODULE. REPLACING THE CABLE RESOLVED THE ISSUE. THESE FAILURES OCCURRED IN SIX BEDS IN A SMALL LONG TERM PEDIATRIC VENTILATOR UNIT. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6075997 ↗
Other Oximeter devices
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy Trilogy 202 Ventilator, United StatesRespironics, Inc. · Oximeter2 deaths1,594 reports
- HAMILTON-G5 159001Hamilton Medical AG · Oximeter18 deaths1,440 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Model EMV+ Vent, Army, MRI 799-EMVP-04Zoll Medical Corporation · Oximeter1 death605 reports
- Trilogy Evo, O2, Japan JP2100X16BRespironics, Inc. · Oximeter4 recalls5 deaths342 reports
- Model EMV+ Vent, Army, COMMERCIAL 799-EMVP-01-01Zoll Medical Corporation · Oximeter1 death497 reports
- HAMILTON-G5 159002Hamilton Medical AG · Oximeter3 deaths451 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page