RD SET™ MD20-12 RD SET Series Patient Cable 4104Recalls & reported problems

Masimo Corporation · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843997010313), received between Dec 9, 2016 and Feb 17, 2017 — classified as 1 death, 2 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message1
  • Inappropriate or Unexpected Reset1
  • Overheating of Device1
  • Sensing Intermittently1

Patient problems

  • Death1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2017Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT " STAFF ARE PULLING OFF CABLE WITHOUT HOLDING THE ACTUAL CONNECTION PIECE ON BOTH THE SENSOR AND CABLE. THE TENSION ON THE CABLE IS CAUSING BREAKDOWN ON THE CABLE". THE PRODUCT WOULD DISPLAY VALUES AND THEN SUDDENLY DISPLAY NO VALUES. NO ERROR MESSAGE OR AUDIBLE ALARM WAS OBSERVED. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6339987 ↗
  • MalfunctionFeb 10, 2017Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    THE CUSTOMER REPORTED THAT THE DEVICE TURNS ON AND OFF. NO CONSEQUENCES OR IMPACT TO PATIENT WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6324472 ↗
  • DeathDec 9, 2016Sensing Intermittently; Device Displays Incorrect Message; Sensing Intermittently; Device Displays Incorrect Message

    IT WAS REPORTED THAT THE CABLE SOMETIMES DOESN'T WORK AND THE MEASURING DEVICE (RAD-8) SHOWS THE ERROR MESSAGE "SEN OFF". THE PATIENT WAS BEING MONITORED AT HOME WITH THE CABLE HAS EXPIRED ON (B)(6) 2016.

    Patient problems: Death

    MAUDE report 6158917 ↗