Skip to content

RAD-G 9895Recalls & reported problems

Masimo Corporation · RAD-G Pulse Oximeter with Temperature · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00843997013789), received between Mar 18, 2024 and Sep 24, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K213676); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 15, 2024Open, Classified

    Their is a potential that Oximeter may automatically power off and on resulting in loss of monitoring.

    Z-1539-2024 · Root cause: Material/Component Contamination · FDA record ↗

Most reported problems

Device problems

  • Unexpected Shutdown5
  • Application Program Freezes or Fails to Launch1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Application Program Freezes or Fails to Launch

    The customer reported, "The only information we have received is that the touchscreen freezes and doesn't work intermittently, It will be like this for a few minutes and then works again. There was also an issue with the rear casing coming away."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50879903 ↗
  • MalfunctionFeb 18, 2026Unexpected Shutdown

    THE CUSTOMER REPORTED THAT THE DEVICE POWERED OFF AND WILL NOT TURN BACK ON. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24384869 ↗
  • MalfunctionDec 27, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-G WAS "BEING A LITTLE TEMPERAMENTAL" AS THE DEVICE "SWITCHES OFF RANDOMLY." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23901427 ↗
  • MalfunctionAug 19, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-GS "WOULD INTERMITTENTLY TURN THEMSELVES OFF WITHOUT ANYONE TOUCHING IT." NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22846659 ↗
  • MalfunctionApr 26, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-G "AUTOMATICALLY TURN[S] ON AND OFF." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19197547 ↗
  • MalfunctionMar 18, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-G DEVICE "TURNING OFF ON INTERMITTENTLY". THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18924214 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K213676. One submission often covers many models, so check the product and lot details in each record.

  • Z-1536-2024 · Feb 15, 2024 · Open, Classified
    Masimo Rad-G, Pulse Oximeter with temperature (W/Sensor), REF:9210, Rx Only,

Other Oximeter devices

All Oximeter devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.