Skip to content

Malyugin Ring System 2.0, 6.25mm MAL-1001-1Recalls & reported problems

Microsurgical Technology, Inc. · Class 1

All models of Malyugin Ring System 2.0, 6.25mm (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Microsurgical Technology Malyugin Ring is an iris retractor used to hold the pupil open during eye surgery. Reports mostly describe the ring twisting, kinking or failing to unfold and deploy correctly from its injector. There are 81 reports that name its identifier, with no deaths and 2 injuries. No recalls name this device, and reports rose from

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
81
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 81 adverse event reports that list this exact device (DI 00844685006120), received between Aug 15, 2025 and Mar 31, 2026 — classified as 2 injuries, 79 malfunctions.

The most frequently reported specific device problem is “Difficult to Fold, Unfold or Collapse” (76 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 5 reportsNov 2025: 11 reportsDec 2025: 10 reportsJan 2026: 5 reportsFeb 2026: 18 reportsMar 2026: 31 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202631
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Fold, Unfold or Collapse76
  • Component Incompatible2
  • Difficult to Insert1
  • Ejection Problem1
  • Retraction Problem1

Patient problems

  • Insufficient Information79
  • Eye Injury2

Recent adverse event reports

The 25 most recent of 81 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Difficult to Fold, Unfold or Collapse

    NEWER RING VERSION WORKS POORLY AND TWISTS ON INSERTION AND HAS FAILED ON 3 CONSECUTIVE PATIENTS. NOT PATIENT INJURY, BUT REQUIRED USE OF IRIS HOOKS WHICH ADDED TO SURGICAL CENTER COST. IRIS HOOKS REQUIRED ON 3 PATIENTS (ONE PATIENT USED TWO RINGS, BOTH FAILED TO WORK).

    Patient problems: Insufficient Information

    MAUDE report 24749701 ↗
  • MalfunctionMar 31, 2026Difficult to Fold, Unfold or Collapse

    RING INJECTED CROOKED. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24748348 ↗
  • MalfunctionMar 31, 2026Difficult to Fold, Unfold or Collapse

    DID NOT DEPLOYED CORRECTLY AND RING DEFECTED. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24747867 ↗
  • MalfunctionMar 31, 2026Difficult to Fold, Unfold or Collapse

    ISSUES WITH THE RING BEING "KINKED" PER THE SURGEON, AND NOT ABLE TO BE INSERTED PROPERLY. IT IS TWO OF THEM. ALL HAVE THE SAME LOT NUMBER.

    Patient problems: Insufficient Information

    MAUDE report 24746576 ↗
  • MalfunctionMar 30, 2026Difficult to Fold, Unfold or Collapse

    DR. (B)(6) ATTEMPTED TO USE TWO DIFFERENT MALYUGIN RINGS FROM THE SAME BOX (INFORMATION ABOVE). HE WAS UNSUCCESSFUL AT DELIVERING THE RING AS IT WAS TWISTED IN THE DELIVERY. THE FRONT RING WOULD DEPLOY BUT THE SECOND TWO ARE TWISTED TOGETHER. HE ATTEMPTED TO UNTWIST IT WITH A SECOND INSTRUMENT BUT WAS UNSUCCESSFUL. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24737209 ↗
  • MalfunctionMar 23, 2026Difficult to Fold, Unfold or Collapse

    MALYUGIN RING DID NOT RETRACK CORRECTLY TO USE FOR IMPLANT. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24670054 ↗
  • MalfunctionMar 23, 2026Difficult to Fold, Unfold or Collapse

    DEFECTIVE, RETRACTED INCORRECTLY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24669939 ↗
  • MalfunctionMar 23, 2026Difficult to Fold, Unfold or Collapse

    MALYUGIN RING MALFUNCTION, NOT ABLE TO FIRE CORRECTLY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24669677 ↗
  • MalfunctionMar 23, 2026Difficult to Fold, Unfold or Collapse

    THE RING WOULD GET TANGLED ON THE HOOK WHEN TRYING TO DEPLOY IT. DOCTOR TRIED 3 TIMES AND ALL OF THEM WERE DEFECTIVE. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24669624 ↗
  • MalfunctionMar 20, 2026Difficult to Fold, Unfold or Collapse

    THE DOCTOR HAD TROUBLE INSERTING 2 OUT OF ONE BOX AND WANTED TO RETURN ALL BOXES WITH THAT LOT NUMBER. ONE DID NOT ADVANCE, ONE CAME OUT OF THE INJECTOR INCORRECTLY, TWO LEADING LOOP WILL COME OUT BUT THE OTHER LOOPS ARE WRAPPED AROUND EACH OTHER. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24656343 ↗
  • MalfunctionMar 17, 2026Difficult to Fold, Unfold or Collapse

    WOULD NOT DEPLOY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24623794 ↗
  • MalfunctionMar 17, 2026Difficult to Fold, Unfold or Collapse

    UPON DEPLOYMENT INTO PATIENT'S EYE, THE RING GOT CAUGHT, "KINKED", CRIMPED IN THE INSERTION DEVICE, RESULTING IN INCOMPLETE DEPLOYMENT. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24622283 ↗
  • MalfunctionMar 16, 2026Difficult to Fold, Unfold or Collapse

    RING CAME OUT TWISTED UPON INJECTION. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24607192 ↗
  • MalfunctionMar 16, 2026Difficult to Fold, Unfold or Collapse

    PER SURGEON, WAS BENT WHEN EJECTED AND WOULD NOT FUNCTION PROPERLY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24607172 ↗
  • MalfunctionMar 16, 2026Difficult to Fold, Unfold or Collapse

    THE DEVICE COILED AND DIDN'T EXPAND PROPERLY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24607146 ↗
  • MalfunctionMar 13, 2026Difficult to Fold, Unfold or Collapse

    WHEN MALYUGIN RING IS DEPLOYED, RING DID NOT COME OUT. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24599452 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    PRODUCT WOULD NOT EXPAND NOR ADVANCE, SUTURE COMES OUT MANGLED. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24553243 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    ONE OF THE RINGS WAS FLAT AND NOT ABLE TO BE MANIPULATED BY THE MD. RING HAD PATIENT CONTACT BUT WAS REMOVED AND PROCEDURE WAS SUCCESSFULLY ABLE TO BE COMPLETED.NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24553163 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    PRODUCT WOULD NOT EXPAND NOR ADVANCE, SUTURE COMES OUT MANGLED. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24552704 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    RINGS FAILED TO EXPAND FULLY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24552352 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    PRODUCT WOULD NOT EXPAND NOR ADVANCE, SUTURE COMES OUT MANGLED. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24552193 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    RETRACT MALYUGIN RING VERY SLOWLY AND NOTICE IT PULLING TO THE LEFT THE EYELETS WERE NOT OVERLAPPING EACH OTHER, SO I STOP AND ASK FOR A NEW ONE. THE SECOND ONE DID NOT ADVANCE PROPERLY IN THE EYE. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24552018 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    RECENTLY HAD TWO DEFECTIVE MALYUGIN RINGS THAT I NEED TO GET REPLACED. THE TWO DEFECTIVE LOT NUMBERS INCLUDE. LUCKILY NO PATIENT INJURY, BUT BOTH HAPPENED UPON DEPLOYMENT INTO THE EYE.

    Patient problems: Insufficient Information

    MAUDE report 24547108 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    RING DID NOT FULLY RETRACT INTO INSERTER AND DEPLOYED BENT/TWISTED, UNUSABLE. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24547054 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    DEVICE DOES NOT DEPLOY CORRECTLY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24546913 ↗

Other Clip, Iris Retractor devices

All Clip, Iris Retractor devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page