Malyugin Ring System 2.0, 7.0mm MAL-1002-1Recalls & reported problems

Microsurgical Technology, Inc. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Microsurgical Technology Malyugin Ring is a small flexible iris ring used to hold the eye open during cataract surgery. The 29 reports mostly describe the ring failing to fold, unfold or open properly — arms locking together, loops twisting or tangling, and kinks — so surgeons had to discard it and open another one. One report describes a tear of the eye's capsule, with 1 injury and no deaths. Reports rose sharply, from 2 to 27 in the past 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00844685006137), received between Oct 15, 2025 and Apr 14, 2026 — classified as 1 injury, 28 malfunctions.

The most frequently reported specific device problem is “Difficult to Fold, Unfold or Collapse” (29 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 7 reportsJan 2026: 2 reportsFeb 2026: 7 reportsMar 2026: 6 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Fold, Unfold or Collapse29

Patient problems

  • Insufficient Information27
  • Capsular Bag Tear1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 14, 2026Difficult to Fold, Unfold or Collapse

    ARMS LOCK ONTO EACH OTHER DURING LOADING AND DON'T DEPLOY. ONE PATIENT CAPSULE TEAR. SUCCESSFUL SURGERY BUT TURNED COMPLEX CHANGED TECHNIQUE.

    Patient problems: Capsular Bag Tear

    MAUDE report 24879617 ↗
  • MalfunctionApr 7, 2026Difficult to Fold, Unfold or Collapse

    A DEFECTIVE RING DURING DEPLOYING INTO PATIENT'S EYE. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24801379 ↗
  • MalfunctionMar 31, 2026Difficult to Fold, Unfold or Collapse

    RING TANGLED BEYOND USE WHEN DEPLOYED FROM DEVICE. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24748839 ↗
  • MalfunctionMar 31, 2026Difficult to Fold, Unfold or Collapse

    THE CUSTOMER FILED A MEDSUN REPORT (B)(4)TO FDA, THE REPORT STATES: THE MALYUGIN RING WAS OPENED FOR USE. HOWEVER, THE SURGEON NOTICED THE RING WAS NOT CORRECTLY ANGLED/DEFECTIVE WHEN DEPLOYED FOR USE. ANOTHER MALYUGIN RING WAS OPENED AND THE SURGEON NOTICED THE SAME ISSUE. PER THE TEAM, MATERIAL OF THE RING WAS CHANGED IN 2.0 VERSION. NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24746020 ↗
  • MalfunctionMar 20, 2026Difficult to Fold, Unfold or Collapse

    SEVERAL OF THE MALYUGIN RINGS FROM THE LOT NUMBERS ABOVE DID NOT DEPLOY CORRECTLY DURING SURGERY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24656329 ↗
  • MalfunctionMar 16, 2026Difficult to Fold, Unfold or Collapse

    RING WOULD NOT DEPLOY PROPERLY. RING WOULD NOT RETRACT PROPERLY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24613221 ↗
  • MalfunctionMar 9, 2026Difficult to Fold, Unfold or Collapse

    THE LOOPS SEEMED TO BE INTERTWINED AND COULD NOT BE USED. THE RING WAS THEN TAKEN OUT AND ANOTHER RING HAD TO BE OPENED. THIS RING DID HAVE PATIENT CONTACT. NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24553021 ↗
  • MalfunctionMar 3, 2026Difficult to Fold, Unfold or Collapse

    DID NOT EXPAND PROPERLY; HAD A KINK. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24503278 ↗
  • MalfunctionFeb 23, 2026Difficult to Fold, Unfold or Collapse

    DR. (B)(6) HAD A 7.0 RING (LOT# 212408) IN WHICH HE SAID THE "PLUNGER OVERRODE THE RING." ANOTHER FAULTY RING TODAY (DR. (B)(6), LOT# 214994, 6.25). SHE STATES IT WAS "STUCK TOGETHER". THERE WAS BRIEF PATIENT CONTACT. NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24421402 ↗
  • MalfunctionFeb 18, 2026Difficult to Fold, Unfold or Collapse

    IN THE PAST COUPLE OF MONTHS, OUR DOCTOR HAS BEEN HAVING ISSUES WITH MAL-1002-1 MISFIRING WHEN INSERTED INTO THE EYE FOR CATARACT SURGERIES. TWO MORE OF MAL-1002-1 THAT MISFIRED. THERE WAS NO HARM TO ANY PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 24385126 ↗
  • MalfunctionFeb 18, 2026Difficult to Fold, Unfold or Collapse

    SCROLLS COMING OUT SIDEWAYS AND MALFORMED ONE NOT RETRACTING. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24384773 ↗
  • MalfunctionFeb 17, 2026Difficult to Fold, Unfold or Collapse

    MALYUGIN RING RETRACTED WITH NO ISSUE INTO INJECTOR, BUT CAME OUT TWISTED WHEN INSERTED INTO THE EYE. NO PATIENT INJURY,

    Patient problems: Insufficient Information

    MAUDE report 24370310 ↗
  • MalfunctionFeb 17, 2026Difficult to Fold, Unfold or Collapse

    MALYUGIN RINGS THAT WERE DEFECTIVE AND DID NOT OPEN PROPERLY UPON INSERTION. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24369660 ↗
  • MalfunctionFeb 2, 2026Difficult to Fold, Unfold or Collapse

    WOULD NOT ADVANCE INTO PATIENTS EYE. INSTRUMENT WAS DISPOSED OF. NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 24240581 ↗
  • MalfunctionFeb 2, 2026Difficult to Fold, Unfold or Collapse

    HE DEVICE DEPLOYED BUT THEN CURLED UP. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24231359 ↗
  • MalfunctionJan 15, 2026Difficult to Fold, Unfold or Collapse

    DEFECTIVE GEOMETRIC CONFIGURATION OF RING ON DEPLOYMENT, RESULTING IN NON-FUNCTIONAL DEVICE. NO PATIENT INJURY. MULTIPLE HAVE SEEN 10-20 TIMES OVER PAST MANY MONTHS IN 2025. LATEST 12/18/25.

    Patient problems: Insufficient Information

    MAUDE report 24090278 ↗
  • MalfunctionJan 15, 2026Difficult to Fold, Unfold or Collapse

    RINGS DO NOT RETRACT WELL AND WHEN THE SURGEON TRIES TO PLACE THE RING, IT EITHER GETS TANGLED OR DOES NOT OPENED CORRECTLY. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24090224 ↗
  • MalfunctionDec 31, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    BOTH WOULD NOT DEPLOY CORRECTLY . HAD [?]PT CONTACT BUT NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23937092 ↗
  • MalfunctionDec 29, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    THE RING MISSHAPEN UPON LEAVING INJECTOR. NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23919123 ↗
  • MalfunctionDec 24, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    DR (B)(6) TRIED TO USE THREE DIFFERENT M RINGS TODAY AND THE RINGS STUCK TOGETHER. THE LOT NUMBER IS 212110 I WOULD LIKE THE WHOLE BOX REPLACED PLEASE. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 23889289 ↗
  • MalfunctionDec 17, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    MALYUGIN RING DID NOT EXIT CARTRIDGE/DEVICE CORRECTLY. PATIENT CONTACT NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23830335 ↗
  • MalfunctionDec 9, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    I HAVE A MALYUGIN RING (MAL-1002-1) THAT CAME OUT OF THE INSERTER DEFORMED. NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23764744 ↗
  • MalfunctionDec 4, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    OPENED THE MALYUGIN RING TO THE SURGICAL FIELD AND IT WOULD NOT LATCH. NO PATIENT INJURY REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 23723657 ↗
  • MalfunctionDec 4, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    THE FIRST WOULD NOT OPEN UP PROPERLY AND WAS UNSHAPED SO THAT IT COULD NOT ATTACH TO IRIS. - PATIENT CONTACT. NO INJURY THE 2ND ONE OF THE RINGS WAS BENT AND WOULD NOT UNFURL. NOTICED BEFORE ENTERING PATIENT EYE. NO PATIENT CONTACTS.

    Patient problems: Insufficient Information

    MAUDE report 23723505 ↗
  • MalfunctionNov 18, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    EXPERIENCED A FAILURE WITH ONE OF THE M. RING 7.0MM INJECTORS FROM OUR RECENT BOX, WHERE THE INJECTOR DID NOT DEPLOY AND INSTEAD CURLED WITHIN THE CAVITY. THIS DISRUPTED THE PROCEDURE, AND WE BELIEVE THE PRODUCT MAY HAVE BEEN FAULTY. MR USED AS PER DFU - SURGEON IS EXPERIENCED. OPENED A NEW MALYUGIN RING (6.25) AS THE COMPLAINT SAMPLE IS THE LAST IN THE BOX. NO PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 23589588 ↗

Other Clip, Iris Retractor devices

All Clip, Iris Retractor devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page