Skip to content

ArcDuo, Guarded Depth Paired Ophthalmic Knife 9mm ARC-0002Recalls & reported problems

Microsurgical Technology, Inc. · The ArcDUO device is a single-use, sterile ophthalmic knife that is intended to make controlled depth paired arcuate incisions on the cornea at or near the limbus. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00844685014316), received between Apr 14, 2026 and Apr 14, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 20, 2025Class IIOpen, Classified

    An ophthalmic knife has a suction set (not direct patient-contacting), which contains springs contaminated with an unidentified residue that may potentially cause eye irritation

    Z-0411-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 14, 2026Use of Device Problem

    PROCEDURES AND PATIENTS WERE SCHEDULED WITH DR. (B)(6) AND TOOK PLACE AT SPECIALTY IN (B)(6). DURING THE SECOND DEMO, SUCTION WAS SUCCESSFUL. DR. (B)(6) PLACED THE BLADES CORRECTLY AND PERFORMED A FIRST TURN. REMOVED THE BLADES AND THE DEVICE. ONLY ONE ARC WAS SUCCESSFUL, UPON WHICH DR. (B)(6) WENT BACK TO PERFORM A SECOND PROCEDURE WITHOUT REMOVING THE FIRST BLADE ON THE SIDE OF THE SUCCESSFUL ARC. AFTER THE SECOND ATTEMPT THE FIRST ARC APPEARED UNCHANGED, AND THE SECOND ARC FAILED TO MATERIALIZE. DR. (B)(6) RESUMED THE CASE AND WENT INTO PHACO. AS FLUIDICS STARTED LEAKING OUT OF THE ARC DR. (B)(6) REALIZED HE HAD PERFORATED THE CORNEA. HE FINISHED HIS PHACO CASE AND SUTURED THE ARC WITH 4 STITCHES.

    Patient problems: Insufficient Information

    MAUDE report 24879883 ↗

Other Knife, Ophthalmic devices

All Knife, Ophthalmic devices →

Sources and updates