Skip to content

MST ArcDUO 9mm, REF: ARC-0002, single-use, sterile… recallZ-0411-2026 · MicroSurgical Technology Inc · initiated Aug 20, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MicroSurgical Technology Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An ophthalmic knife has a suction set (not direct patient-contacting), which contains springs contaminated with an unidentified residue that may potentially cause eye irritation

FDA root cause category: Material/Component Contamination

What the company is doing

On 8/20/2025, a recall notice was sent to a surgery center that was asked to return the affected devices and provide information on affected device disposition. The customer was also asked, if they further distributed affected devices, to notify their customers of the recall For additional information, contact the firm at [email protected]

Recall details

Recall number
Z-0411-2026
Status
Open, Classified
Initiated by the company
Aug 20, 2025
Posted by the FDA
Oct 29, 2025
Recalling firm
MicroSurgical Technology Inc
Quantity
8
Product code
HNN
Distribution
US Nationwide distribution in the state of GA.

Product description

MST ArcDUO 9mm, REF: ARC-0002, single-use, sterile ophthalmic knife, used with Suction Set, Part: 48-1000-01

Code information (models, lots, serial numbers)

UDI:00844685014316. Lots: 207552 and 208238

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page