GEM COUPLER Microvascular Anastomotic Instrument GEM2740Recalls & reported problems

Synovis Micro Companies Alliance, Inc · A reusable Instrument designed to hold the GEM COUPLER device and approximate the GEM COUPLER rings during vessel anastomosis · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00844735005233), received between Apr 21, 2026 and Apr 21, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Mechanical Jam” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam2

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Failure to Anastomose2
  • Hemorrhage/Blood Loss/Bleeding2
  • Pneumothorax2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 21, 2026Mechanical Jam

    IT WAS REPORTED THAT A GEM COUPLER MICROVASCULAR ANASTOMOTIC INSTRUMENT PRESENTED AN ABNORMAL RESISTANCE ("SLIGHT SNAGGING") PRIOR TO BREAST RECONSTRUCTION SURGERY. THE DEVICE WAS RETURNED FOR LUBRICATION AND RESTERILIZATION, AND A SECOND VEIN COUPLER WAS OPENED AS A BACKUP; BOTH WERE AVAILABLE FOR USE PRIOR TO SURGERY. DURING VEIN TO VEIN ANASTOMOSIS, NEITHER INSTRUMENT FUNCTIONED AS INTENDED, AS THE USER WAS UNABLE TO CONNECT THEM TO THE DISPOSABLE COUPLER COMPONENTS. THE USER REMARKED THAT THERE WAS "SOMETHING WRONG WITH THE INTERNAL SPRING MECHANISM", AND THEY COULD VISUALLY OBSERVE THAT SOMETHING WAS OBSTRUCTING IT. MULTIPLE DISPOSABLE COMPONENTS WERE ATTEMPTED WITHOUT SUCCESS. AN OLDER COUPLER INSTRUMENT (ANASTOMOTIC INSTRUMENT) WAS SUBSEQUENTLY USED BUT DEMONSTRATED SIGNIFICANT RESISTANCE. ALTHOUGH AN ANASTOMOSIS WAS ACHIEVED, AN ANASTOMOTIC LEAK WAS DETECTED, REQUIRING A NEW ANASTOMOSIS. DUE TO VESSEL SHORTENING, ADDITIONAL SURGICAL INTERVENTION WAS REQUIRED. PART OF THE RIB HAD TO BE RESECTED, DURING WHICH A CUT OCCURRED AND SIGNIFICANT HEMORRHAGE FROM THE INTERNAL MAMMARY ARTERY HAPPENED. THE SURGEON CLAMPED THE "BLEEDING" PREVENTING ANY MAJOR BLOOD LOSS. DURING SUCH COMP

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding; Pneumothorax

    MAUDE report 24947629 ↗
  • InjuryApr 21, 2026Mechanical Jam

    IT WAS REPORTED THAT A GEM COUPLER MICROVASCULAR ANASTOMOTIC INSTRUMENT PRESENTED AN ABNORMAL RESISTANCE ("SLIGHT SNAGGING") PRIOR TO BREAST RECONSTRUCTION SURGERY. THE DEVICE WAS RETURNED FOR LUBRICATION AND RESTERILIZATION, AND A SECOND VEIN COUPLER WAS OPENED AS A BACKUP; BOTH WERE AVAILABLE FOR USE PRIOR TO SURGERY. DURING VEIN TO VEIN ANASTOMOSIS, NEITHER INSTRUMENT FUNCTIONED AS INTENDED, AS THE USER WAS UNABLE TO CONNECT THEM TO THE DISPOSABLE COUPLER COMPONENTS. THE USER REMARKED THAT THERE WAS "SOMETHING WRONG WITH THE INTERNAL SPRING MECHANISM", AND THEY COULD VISUALLY OBSERVE THAT SOMETHING WAS OBSTRUCTING IT. MULTIPLE DISPOSABLE COMPONENTS WERE ATTEMPTED WITHOUT SUCCESS. AN OLDER COUPLER INSTRUMENT (ANASTOMOTIC INSTRUMENT) WAS SUBSEQUENTLY USED BUT DEMONSTRATED SIGNIFICANT RESISTANCE. ALTHOUGH AN ANASTOMOSIS WAS ACHIEVED, AN ANASTOMOTIC LEAK WAS DETECTED, REQUIRING A NEW ANASTOMOSIS. DUE TO VESSEL SHORTENING, ADDITIONAL SURGICAL INTERVENTION WAS REQUIRED. PART OF THE RIB HAD TO BE RESECTED, DURING WHICH A CUT OCCURRED AND SIGNIFICANT HEMORRHAGE FROM THE INTERNAL MAMMARY ARTERY HAPPENED. THE SURGEON CLAMPED THE "BLEEDING" PREVENTING ANY MAJOR BLOOD LOSS. DURING SUCH COMP

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure to Anastomose; Hemorrhage/Blood Loss/Bleeding; Pneumothorax

    MAUDE report 24947595 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →