Skip to content

Solanas 63035-16Recalls & reported problems

Alphatec Spine, Inc. · POLYAXIAL POSTERIOR THORACIC SCREW, Ø3.5mm X 16mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00844856006607), received between Dec 13, 2016 and Dec 13, 2016 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Break1
  • Device Dislodged or Dislocated1
  • Fracture1

Patient problems

  • Failure of Implant1
  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 13, 2016Break; Fracture; Device Dislodged or Dislocated

    AN INTERNATIONAL CUSTOMER ((B)(6)) REPORTED THAT A YOUNG MALE PATIENT WITH GOOD BONE QUALITY UNDERWENT REVISION SURGERY TO REMOVE AND REPLACE 3 BROKE SOLANAS SCREWS ON (B)(6) 2016. THE SCREWS FRACTURED BROKE MID-WAY OF THE THREADED SHAFT AT THE C6 AND C7. THE PROXIMAL SECTIONS OF THE SCREWS WERE REMOVED WITHOUT ISSUE, WHILE THE DISTAL THREADED PORTIONS REMAIN ENTRAPPED WITHIN THE PATIENTS C6/C7 CERVICAL VERTEBRAE'S. THE SOLANAS FIXATION SYSTEM WAS ORIGINALLY IMPLANTED ON (B)(6) 2016.

    Patient problems: Failure of Implant; Foreign Body In Patient

    MAUDE report 6170634 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page