Skip to content

Illico 73724Recalls & reported problems

Alphatec Spine, Inc. · SCREW EXTENDER · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00844856041691), received between May 18, 2016 and Apr 17, 2024 — classified as 2 injuries, 11 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break9
  • Fracture3
  • Physical Property Issue2
  • Detachment of Device or Device Component1

Patient problems

  • No Known Impact Or Consequence To Patient9
  • No Information3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2024Break; Break

    IT WAS REPORTED THAT THE SCREW EXTENDER WAS FOUND BROKEN AFTER A WASH CYCLE. NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19125935 ↗
  • MalfunctionMar 5, 2020Break

    LOCKING TAB SNAPPED OFF AFTER ROD WAS FULLY SEATED WITHIN TULIP OF SCREW.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9792719 ↗
  • MalfunctionJul 17, 2018Detachment of Device or Device Component

    AN INTERNATIONAL CUSTOMER ((B)(6)) REPORTED THAT THE SHRAPNEL SNAPPED OFF DURING THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7695191 ↗
  • MalfunctionJul 17, 2018Break

    AN INTERNATIONAL CUSTOMER ((B)(4)) REPORTED ONE SIDE OF SHRAPNEL WAS BROKEN DURING THE OPERATION.

    Patient problems: No Information

    MAUDE report 7695144 ↗
  • MalfunctionJul 17, 2018Break

    AN INTERNATIONAL CUSTOMER ((B)(4)) REPORTED THE SHRAPNEL IS BROKEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7695018 ↗
  • MalfunctionJul 17, 2018Break

    AN INTERNATIONAL CUSTOMER ((B)(6)) REPORTED THE SHRAPNEL IS BROKEN.

    Patient problems: No Information

    MAUDE report 7694734 ↗
  • MalfunctionMay 15, 2017Break

    AN INTERNATIONAL CUSTOMER ((B)(6)) REPORTED DURING THE PROCEDURE, THE DEVICE IS USED TO HALF THE TIME, THE SHRAPNEL IS BROKEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6567184 ↗
  • InjuryApr 28, 2017Break

    AN INTERNATIONAL CUSTOMER ((B)(6)) REPORTED THE BLADE OF AN ILLICO SCREW EXTENDER IS DAMAGED. THE EVENT CAUSED OVER A 30 MINUTE DELAY IN THE CASE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6530165 ↗
  • InjuryApr 28, 2017Break

    AN INTERNATIONAL CUSTOMER (TAIWAN) REPORTED THE BLADES OF TWO (2) ILLICO SCREW EXTENDERS ARE DAMAGED. THE EVENT CAUSED OVER A 30 MINUTE DELAY IN THE CASE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6529963 ↗
  • MalfunctionDec 7, 2016Break

    A 73724 INSTRUMENT BROKEN.

    Patient problems: No Information

    MAUDE report 6152085 ↗
  • MalfunctionNov 21, 2016Fracture; Physical Property Issue

    AN INTERNATIONAL CUSTOMER ((B)(6) REPORTED THAT THE SIDE CLIP OF THE SCREW EXTENDER WAS FRACTURED. IN THIS CASE, SURGEON HAD USED THE SCREW EXTENDER ALIGNMENT GUIDE TO ATTACH THE EXTENDERS ON IMPLANTS. AND SURGEON ALSO HAD USED THE SCREW EXTENDER REMOVER TO REMOVE THE SCREW EXTENDER AT THE END OF THE SURGERY. BUT THIS EXTENDER STILL WAS DAMAGED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6118258 ↗
  • MalfunctionNov 21, 2016Fracture; Physical Property Issue

    AN INTERNATIONAL CUSTOMER ((B)(6)) REPORTED THAT THE SIDE CLIP OF THE SCREW EXTENDER WAS FRACTURED. IN THIS CASE, SURGEON HAD USED THE SCREW EXTENDER ALIGNMENT GUIDE TO ATTACH THE EXTENDERS ON IMPLANTS. AND SURGEON ALSO HAD USED THE SCREW EXTENDER REMOVER TO REMOVE THE SCREW EXTENDER AT THE END OF THE SURGERY. BUT THIS EXTENDER STILL WAS DAMAGED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6118217 ↗
  • MalfunctionMay 18, 2016Fracture

    AN INTERNATIONAL CUSTOMER (B)(6) REPORTED THAT THE SIDE CLIP OF THE SCREW EXTENDER HAD FRACTURED. IN THIS CASE, THE SURGEON HAD USED THE SCREW EXTENDER ALIGNMENT GUIDE TO ATTACH THE EXTENDERS TO THE IMPLANTS. THE SURGEON ALSO HAD USED THE SCREW EXTENDER REMOVER TOOL TO DETACH THE SCREW EXTENDER AT THE END OF THE SURGERY, BUT THE EXTENDER STILL WAS DAMAGED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5663266 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page