Skip to content

Illico 73875-40Recalls & reported problems

Alphatec Spine, Inc. · Ti CANNULATED POLYAXIAL SCREW 7.5mm X 40mm · Class 2 · Implant

All models of Illico (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00844856043329), received between Sep 9, 2019 and Oct 8, 2020 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K123623); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component1
  • Noise, Audible1

Patient problems

  • Failure of Implant1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 8, 2020Break; Detachment of Device or Device Component

    AFTER THE FINAL TIGHTENING, THE SURGEON FOUND THAT THE SET SCREW POPPED OFF BY IMAGING. THE SURGEON RETRIEVED THE SET SCREW AND TRIED TO RE-INSTALL IT BUT COULD NOT. THE IMAGES ALSO CONFIRMED THAT THE POLYAXIAL SCREW HEAD/TULIP WAS BROKEN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10651167 ↗
  • InjurySep 9, 2019Break; Noise, Audible

    THE SURGEON TRIED TO REMOVE THE SCREW EXTENDER FROM THE SCREW USING THE RECOMMENDED REMOVER TOOL, BUT WAS UNSUCCESSFUL. HE THEN ATTEMPTED REMOVAL VIA THE REMOVAL AID TOOL AND HEARD A BREAKING SOUND. HE FOUND THAT THE SCREW HEAD/TULIP HAD BROKEN. REMOVED AND REPLACEMENT OF THE SCREW CAUSED AN HOUR DELAY IN THE CASE.

    Patient problems: Failure of Implant

    MAUDE report 8974885 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123623. One submission often covers many models, so check the product and lot details in each record.

  • Z-2086-2013 · Jun 28, 2013 · Terminated
    Illico Ti Cannulated Polyaxial Screw 7.5mm X 40mm. The Illico MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by p

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page