Arsenal 47027Recalls & reported problems

Alphatec Spine, Inc. · SET SCREW · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00844856070325), received between Feb 24, 2016 and Apr 25, 2017 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Detachment Of Device Component1

Patient problems

  • No Code Available2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2017Device Dislodged or Dislocated

    AT 4.5 MONTH POST-OP X-RAYS REVEALED THE ARSENAL SET SCREW LOCATED AT THE S1 HAD DISENGAGED/BACKED OUT FROM THE HEAD OF THE POLYAXIAL SCREW BODY. HOWEVER, THE ROD REMAINED PROPERLY POSITIONED AND ALLOWED FUSION/HEALING TO OCCUR. THE CONSTRUCT HAS BEEN IMPLANTED FOR APPROXIMATELY 2.5 YEARS WITHOUT ISSUE. THERE ARE NO PLANS FOR REVISION AT THIS TIME. THE ARSENAL FIXATION SYSTEM WAS ORIGINALLY IMPLANTED ON (B)(6) 2014.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6518796 ↗
  • InjuryOct 11, 2016Detachment Of Device Component

    TWO ARSENAL SET SCREWS (47027) WERE FOUND DETACHED FROM THE TULIP OF THE MATING POLYAXIAL SCREWS LOCATED AT BOTH THE RIGHT AND LEFT SIDES OF THE S1. THE PATIENT WAS REPORTED TO HAVE ALREADY FUSED/HEALED. REVISION SURGERY WAS CONDUCTED ON (B)(6) 2016 TO REMOVE THE L4-S1 CONSTRUCT.

    Patient problems: No Code Available

    MAUDE report 6017885 ↗
  • InjuryFeb 24, 2016Device Dislodged or Dislocated

    REVISION SURGERY WAS CONDUCTED ON (B)(6) 2016 TO REMOVE AND REPLACE 3 ARSENAL SET SCREWS AT THE L5-S1. THE SET SCREWS WERE FOUND TO HAVE BACKED OUT OF POSITION ONE MONTH POST-OP. THE ARSENAL SPINAL FIXATION WAS ORIGINALLY IMPLANTED ON (B)(6) 2015.

    Patient problems: No Code Available

    MAUDE report 5457228 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices